Research Article Volume 9 Issue 4
1Gynecology and Obstetrics, Mexico
2Gynecology and Obstetrics, Gynecology and Obstetrics Service, Silao General Hospital, Mexico
3Gynecology and Obstetrics, Head of Reproductive Health, General Hospital, Mexico
4Medical Epidemiologist, In charge of the Epidemiology Service of the Silao General Hospital, Mexico
Correspondence: Dr. Eduardo de la Luna and Olsen, Gynecology and Obstetrics, Cerro del Erizo # 39, Col. Valenciana, CP 36240. Guanajuato, Gto, Mexico, Tel (01) 473 73 38068
Received: May 16, 2018 | Published: August 2, 2018
Citation: Olsen EL, Sánchez BC, López LN, et al. Outcome of the use of bakri balloon in obstetric hemorrhage. Obstet Gynecol Int J. 2018;9(4):281-286. DOI: 10.15406/ogij.2018.09.00348
Objective: To present the results obtained by using the Bakri Balloon to obstetric control hemorrhage.
Material and method: Prospective, observational study within two inpatient medical care units since January 1 to December 31 2016. Twenty patients with post-partum and trans cesarean hemorrhage. All of them were applied the Bakri Balloon because of failure to respond to uterotonic drug therapy. The following were analyzed:
Results: The amount of bleeding after insertion, both in postpartum and trans cesarean was reduced and there was an adequate response in hemoglobin levels. The average time between hemorrhage diagnosis and balloon insertion was thirty minutes; average insufflation time, five minutes, and average volume supplied 400mL. Balloon's average use time, 29.5hours. In 95% of the cases there was a positive response for hemorrhage control, with no complications derived from use.
Conclusions: The Bakri Balloon proved to be a useful, quick and complication -free therapy for controlling obstetric hemorrhage.
Keywords: obstetric hemorrhage, Bakri Balloon
Obstetric hemorrhage has increased its frequency1–3 for different reasons, among which are the increase in cesarean section and inadequate use of uterotonics, among others.4–6 It occupies the second cause of maternal deaths in our country and in some States, the first.7–9 Several strategies have been carried out for its reduction, without obtaining the expected results.3,8,9 Several decades ago the most practiced surgical treatment, once failed ls medical measures, was obstetric hysterectomy, which besides being a limiting fertility procedure in many patients, and was not without many complications.10,11 Nowadays, this behavior is increasingly questioned as the first therapeutic procedure and even in many countries it is the subject of medical litigation.12,13 Others, though effective, surgical techniques such as hypogastric artery ligation or devascularization of the uterine vessels require significant medical training and good experience, both often absent in our hospital means.14–16 The advent of procedures considered more conservative, such as the application of the Bakri intrauterine balloon, which requires a low and rapid learning curve, has demonstrated its usefulness for the effective control of obstetric hemorrhage and with minimal complications.17,18 Up to this date there are multiple reports on their use.9,20,21
The objective of this report is to present the results obtained in two hospitals with the use of Bakri's intrauterine balloon to control obstetric hemorrhage both in the postpartum and in the trans-cesarean, its relation to the amount of hemorrhage and hemoglobin levels before and after its application.
Prospective, longitudinal study performed in two hospital units: Hospital Silao General of the Ministry of Health of the State of Guanajuato and General Hospital Los Reyes, Secretary of Health of the State of Michoacan, in a span of 1. From January to December 31, 2016, of consecutive cases that went to the medical units in the period indicated. We included 20 patients who presented both postpartum and trans cesarean hemorrhage. Hemorrhage was considered from 500mL22 and / or when the patient presented data on hypovolemia, characterized by hypotension (equal to or less than 90/60 for systolic / diastolic), tachycardia (equal to or greater than 100). l/min), tachypnea (equal to or greater than 20r/min), neurological changes (confusion, restlessness, decreased alertness), decreased urine output (less than 20mL/hour) and decreased capillary refill.4,22 The amount of bleeding was calculated in two ways: 1st) by gravimetry, weighing with gauze and compresses during the procedures and 2nd) with collection bag type Brass -V, Calibrated obstetric drape (MaternovaTM). All the patients were initially treated with uterotonics. First choice oxytocin was used at a rate of 10U IV, diluted in 10 ml of Hartman solution passing slowly (1min), if no response was obtained; Carbetocin was continuedTM (Lab. Ferring) 100mcgs, 1ml IV in 1minute. In all patients the hemoglobin admission unit determined, a post-hemorrhagic event twenty - four hours time after use with the ball or before removal thereof.23,24
All patients without response to medical treatment were given an intrauterine Bakri balloon (Cook Medical, Bloomington, IN. USA, Demesa, Mexico City) (Figure 1).
The application of the balloon in postpartum cases was performed with the following technique:
Visualization of the cervix through vaginal valves, asepsis and antisepsis of the region, clamping of both cervical lips. The balloon was applied by guide with ring clamp to the uterine cavity. The balloon was filled with a 60mL syringe, with sterile, warm saline. The amount to be infused depended on the response to control the bleeding, without exceeding 500mL total. The filling speed was recorded. The correct placement of the balloon was corroborated, clinically in the initial phase and a gauze plug was placed in the bottom of the bag to favor the anterior position towards the pubis of the cervix and thus avoid the possible externalization of the balloon. The insufflation catheter was fixed to the thigh of the patient and connected to a collection bag to obtain a strict control of the amount of bleeding. At the end of the procedure, the correct placement of the balloon by abdominal ultrasound was corroborated.
In cases of trans cesareal balloon application, the technique was performed through hysterotomy, directing the balloon to the fund us of the uterus and the catheter for filling was passed trans cervical until vagina where a helper led him to the outside and proceeded to fill it with the same technique and quantity of solution required and described for postpartum cases. The hysterorraphy was performed with the usual technique. At the end of the procedure, its proper placement by abdominal ultrasound was also corroborated. In all cases the ball was placed by different doctors Gineco -Obstetras with training in the procedure.
L as patients remained completely still during the time the ball had been prescribed prophylactic antibiotic, acinus Ciproflox 1g intra venous (Basic table and catalog of medications, 2014. key 1937.00). All patients had hemoglobin determination and coagulation tests on admission to the care units. In to dorslors cases determination of hemoglobin and coagulation tests was performed one hour after the hemorrhage and 24hours after treatment with blood components or before removing or d the ball.
T he result is divided into:
Positive response: Control of bleeding no greater than 150-200mL March 2 in 24hours and recovered ban signs of hypovolemia. In these cases the balloon was removed between 24 to 48hours
Negative response: When control of bleeding was not obtained and hypovolemia data persisted. In these cases, another therapeutic measure was taken with the immediate removal of the balloon.
S and performed the following analysis:
Descriptive statistics of numerical (mean and standard deviation) and categorical (proportions) data were used. Due to the small size of the sample, since the distribution of the data was not normal, non-parametric statistics were used. For the comparison of numerical variables before and after the use of the Bakri balloon in patients with uterine bleeding, the Wilcoxon test of the signed ranges was used. To compare numerical data of the patients with cesarean section and delivery, the amount of bleeding comparing them and the hemoglobin values at the different times indicated, the Mann Whitney U test was used. A value of p <0.05 was considered statistically significant. For the analysis of the data, the IBM SPSS version 22 statistical package was used
Twenty patients were treated during the study period. According to the seven points poured into the material and method, the following results were obtained:
No patients |
Age |
AUG |
Weeks Gestation |
Diagnosis |
Pretreatment ball Bakri |
1 |
21 |
G1, C1 |
36 |
Previous placenta |
Ox, Mi, Ca |
2 |
21 |
G2, C2 |
40.2 |
TP Extended |
Ox, Mi, Ca |
18 hours |
|||||
3 |
18 |
G1, P1 |
38 |
Uterine hypotonia |
Ox, Ca |
4 |
19 |
G2, P2 |
37 |
Uterine hypotonia |
Ox, Mi, Ca |
5 |
21 |
G2, P1, C1 |
40 |
Uterine hypotonia |
Ox, Mi, Ca |
6 |
20 |
G2, P2 |
38.5 |
Uterine hypotonia |
Ox, Mii, Ca |
7 |
24 |
G3, P3 |
41 |
Uterine hypotonia |
Ox, Mi, Ca |
8 |
17 |
G2, P2 |
40 |
Uterine hypotonia |
Ox, Mi, Ca |
9 |
32 |
G3, C2, A1 |
39.1 |
Previous placenta |
Ox, Mi, Ca |
10 |
2. 3 |
G3, P1, C1, A1 |
32.6 |
RPM. Prolonged TP 16 hours |
Ox, Ca |
11 |
19 |
G1, P1 |
38.6 |
Prolonged TP 15 hrs. macrosomia |
Ox, Mi, Ca |
12 |
3. 4 |
G5, P4, A1 |
39.1 |
Uterine hypotonia |
Ox, Mi, Ca |
13 |
42 |
G6, P5, A1 |
30 |
Previous placenta |
Ox, Mi, C |
14 |
35 |
G4, P4 |
39.6 |
Fortuitous birth |
Ox, Mi, Ca |
Uterine hypotonia |
|||||
15 |
2. 3 |
G2, P1, C1 |
39.6 |
severe pre-eclampsia |
Ox, Mi, Ca |
16 |
32 |
G1, P1 |
39.4 |
severe pre-eclampsia |
Ox, Mi, Ca |
17 |
18 |
G1, P1 |
41.3 |
Prolonged TP 19 hrs |
Ox, Mi, Ca |
Uterine hypotonia |
|||||
18 |
20 |
G1, P1 |
39.4 |
Uterine hypotonia |
Ox, Mi, Ca |
19 |
22 |
G2, P1, A1 |
39 |
Uterine hypotonia |
Ox, Mi |
20 |
31 |
G2, P2 |
36 |
RPM, prolonged PT 17 hours |
My, Ca |
Table 1 Clinical characteristics. Diagnosis and initial treatment
G, Gesta; P, Child birth; C, Cesarean; A, Abortion; TP, Labor; RPM, premature rupture of membranes; Ox, Oxytocin; Mi, Misoprostol; Ca, Carbetocin
No. Patient |
Total amount of bleeding (mL) |
Time between diagnosis and placement (min) |
Time of filling (min) |
Amount of soloution used (mL) |
Time of permanence (hours) |
1 |
2300 |
20 |
5 |
300 |
48 |
2 |
1800 |
20 |
5 |
400 |
24 |
3 |
1200 |
15 |
6 |
500 |
48 |
4 |
3500 |
5 |
7 |
500 |
30 |
5 |
1200 |
5 |
5 |
450 |
26 |
6 |
1200 |
10 |
5 |
500 |
24 |
7 |
900 |
5 |
5 |
350 |
48 |
8 |
1800 |
20 |
5 |
340 |
24 |
9 |
2300 |
5 |
5 |
500 |
48 |
10 |
1300 |
5 |
4 |
400 |
24 |
11 |
1000 |
5 |
5 |
500 |
24 |
12 |
800 |
20 |
5 |
300 |
24 |
13 |
2600 |
5 |
5 |
300 |
48 |
14 |
1000 |
40 |
10 |
350 |
24 |
15 |
2000 |
5 |
5 |
500 |
24 |
16 |
1500 |
5 |
5 |
400 |
24 |
17 |
1500 |
5 |
3 |
300 |
24 |
18 |
1500 |
5 |
5 |
500 |
48 |
19 |
1000 |
5 |
5 |
250 |
24 |
20 |
1000 |
20 |
5 |
350 |
24 |
Table 2 Amount of bleeding, time between diagnosis and placement of the balloon, filling time, amount of solution used and dwell time
* 1) By gravimetry and 2) with collection bag Brass-V, Calibrated obstetric drape (Maternova TM) type.
Bleeding before application of the ball * |
Bleeding after the application of the ball ** |
||
Postpartum mL |
Postpartum mL |
After cesarean section mL |
Cesarean section mL |
1407±672 |
2060±513 |
128±123 |
140±82.1 |
Table 3 Amount of bleeding between cases of delivery and cesarean section. Before and after the application of the ball.
* p = 0.03 ** p = 0.001
Hemoglobin to the Entrance g/dL* |
Hemoglobin 1 hour Postsangrado g/dL* |
Post- treatment hemoglobin g/dL** |
11.8 |
8.1 |
9.6 |
Table 4 Mean levels of hemoglobin on admission, one hour post- graduated and later to treatment with blood components and removal of the balloon
* p = 0.008 ** p = 0.002
In only one case (5%), a positive response with the ball was not obtained. This patient presented a quantified bleeding in 3000mL, secondary to a uterine hypotonia. After the placement of the balloon at 30minutes, the loss was 500mL, so the balloon was removed. The patient was transferred to intensive care, where the presence of coagulopathy due to consumption was corroborated. After volume restitution, fresh plasma and platelet concentrate with improvement in coagulation times, total abdominal hysterectomy was performed. There was no maternal death in the 20 cases studied.
There were no complications in the 19 patients who successfully used the Bakri balloon.
The successes of Bakri's balloon in obstetric hemorrhage range from 100% 18, 24 to 80%.25 The variation in their results may depend on the cause of the bleeding, which is multiple.18,25–27 In this report predominated uterine hypotonia first cause and prolonged labor as second causes that seem to have a better response with the use of intrauterine balls.28,29 In both cases apparently the pathophysiology may be due to difficulty of the muscle fiber to occlude the uterine vessels in the immediate postpartum period.28,30 Bakri's balloon, by producing a hydrostatic effect on the vessels and occluding them, favors its haemostasis17,18,28 as well as activating hemostatic mechanisms and favoring the formation of a stable clot.31 Its use in other pathologies, such as placental accreta, reduces its effect.26,32
One issue discussed with its use is the amount of solution needed to fill the balloon.24 In our setting, Ortega-Vadillo correlated the volume to be infused with gestational age and the weight of the newborn, with good results, 94.2% of success.32 In our report the criterion for the amount of the filling was the cessation of bleeding, in quantities that were 250 to 500cc, with an average of 400cc and we observed similar results. Therefore, we consider that any of these methods for filling the balloon provides similar results.
There are currently no reports in the literature that correlate the use of the Bakri balloon in cases of delivery and caesarean section, before and after its application in relation to hemoglobin levels, also before and after its use. In our report we found a statistically significant difference with the use of the balloon, in both cases. Therefore, we cannot compare our results with other studies. More studies are required comparing these variables.33
In the present report, there was no complication with the use of the ball. Some authors have reported accidents such as inadvertent uterine rupture, perforation 33 or spontaneous expulsion.34 The use of tamponade in the vaginal fund us to elevate the cervix towards the retro pubis, the practice of ultrasound to corroborate its adequate position and to keep the patient in absolute rest during the time of use, seem to be favorable measures to avoid this type of complications.
The success of the Bakri balloon in our report, 95%, is similar to multiple studies already reported.18,20,21,24,32 Its ease of application, being a conservative, simple procedure with a fast and easily replicable learning curve make it suitable for the control of obstetric hemorrhage. In our report, the average time for its application was 5minutes, a time that is not observed with other procedures that are equally effective, but more complex.
Bakri's balloon showed a favorable response for its use in obstetric hemorrhage. An adequate control of it is obtained, observing a significant decrease in the amount of bleeding, both in postpartum and trans cesarean cases, recovering the hemoglobin values before and after its use. Its level of security is adequate, easy to apply and with simple training. It has a successful result rate, in this report of 95%, very similar to other reports in the literature.
None.
The author declares that they do not have any conflicts of interest.
©2018 Olsen, et al. This is an open access article distributed under the terms of the, which permits unrestricted use, distribution, and build upon your work non-commercially.