Submit manuscript...
International Journal of
eISSN: 2475-5559

Petrochemical Science & Engineering

Research Article Volume 2 Issue 6

Measuring Suspended Particle Size with High Accuracy

Huili Zhang,1 Jan Baeyens,1,2 Qian Kang3

1Beijing University of Chemical Technology, College of Life Science and Technology, China
1Beijing University of Chemical Technology, College of Life Science and Technology, China
2University of Warwick, School of Engineering, Coventry, UK
2University of Warwick, School of Engineering, Coventry, UK
3KU Leuven, Department of Chemical Engineering, Belgium
3KU Leuven, Department of Chemical Engineering, Belgium

Correspondence: Qian Kang, KU Leuven, Department of Chemical Engineering, Belgium

Received: July 11, 2017 | Published: August 9, 2017

Citation: Zhang H, Baeyens J, Kang. Measuring suspended particle size with high accuracy. Int J Petrochem Sci Eng. 2017;2(6):201-206. DOI: 10.15406/ipcse.2017.02.00058

Download PDF

Abstract

The average particle size and its distribution are major characteristics of a powder. Instrumental techniques, i.e. particle size analysers, are frequently used and span a wide size range. The selection of the most suitable measurement procedure for a particular application may present a problem, aggravated by the complexity of comparing results produced by different methods.

The laser diffraction method is the most popular method of analysis. The authors therefore investigated the conditions required to improve the accuracy of the measurement. These conditions include the sample preparation (solvent, dispersant), the dispersion itself, and the optical characteristics of the particles. The paper summarizes the findings and defines general rules to be respected.

Keywords: Powders; Particle Measuring Methods; Particle Size Analysis; Particle Size Distribution Analysis; Laser Diffraction

Introduction

The average particle size and its distribution are undoubtedly major characteristics of a powder, and the average particle size occurs in all correlations related to the powder behaviour in, e.g., packed, moving or fluidized beds, in pneumatic conveying and in dust filtration. When traditional sieving can no longer be used for smaller particles, instrumental techniques, i.e. particle size analysers, are used. These analysers extended the measured particle size to the micron or sub-micron sizes, and span a fairly wide size range. Among these instrumental techniques, major developments have taken place, resulting in a wide range of equipment and including, e.g., sedimentation, electrical sensing methods, laser diffraction, dynamic light scattering (for molecules and sub-micron particles), X-ray scattering, acoustic methods, and/or focused beam techniques.1-15 The common equipment and a non-exhaustive list of companies involved are given in Table 1.

Method

Company

 

Laser diffraction & scattering

SYMPATEC
MALVERN
COULTER
HORIBA
LEEDS & NORTHRUP
CILAS
SHIMADZU

www.sympatec.com
www.malvern.com
www.beckmancoulter.com
www.horiba.com
www.leedsandnorthrup.com
www.cilas.com
www.shimadzu.com

Photo sedimentation

HORIBA
SHIMADZU
BROOKHAVEN

www.horiba.com
www.shimadzu.com
www.bic.com

X-ray sedimentation

MICROMERITICS

www.micromeritics.com

Light obscuration

GALAI INSTRUMENTS

www.cortera.com

Electrical sensing zone

COULTER
MICROMERITICS

www.beckmancoulter.com
www.micromeritics.com

Dynamic light scattering

MALVERN
BROOKHAVEN

www.malvern.com
www.bic.com

Image analysis

ALLIED VISION

www.alliedvisiontec.com

Acoustic methods
Focused beam

SYMPATEC

www.sympatec.com

Table 1 Common particle size analysers and some of the suppliers.

The final objective of the particle size analysis technology is to accurately measure the particle size distribution in any size range. And yet, repeat determinations often lack consistency and accuracy. It is essential to respect a number of rules, related to the sample preparation, the measurement procedure and to the analysers themselves. Procedures should hence be standardized and this has been presented in the ISO standards for each of the measurement techniques.16-21 Each standard describes at some degree the advantages and limitations of particular methods. Unfortunately, no information is given about comparing results produced by different methods, and such a comparison often creates the impression that results are very different.

These differences are due to several reasons, associated with:

  1. Sample preparation and handling
  2. Breaching instrument specifications
  3. Differences in presenting the particle size distributions.

The laser diffraction and scattering method is one of the most popular particle size analysers. The authors therefore conducted a number of experiments on a Malvern laser analyser to establish conditions and associated operating conditions required to improve the accuracy of the measurement.

The accuracy of the particle size analysis is largely dependent on a number of parameters, including the sample preparation, the degree of dispersion and the analysis itself. In most common suspension analysers, particles are dispersed in a solvent, mostly using a dispersant to lower the surface tension.

The main parameters that are to be considered in particle size measurement are listed in Table 2. The effect of these parameters on the particle sizing results will be illustrated in Section 3 of the paper.

Parameter

Effect

Comments

Cohesion

Agglomeration
(increase in effective dp)

Mostly Van der Waals forces; Counteract by promoting wetting and rupture

Wettability

Cover particle surface with liquid; Replace air from external/internal surface

Liquid should be less polarisable than solids; Use surfactants

Rupture of agglomerates

Make particles individual

Induce shear on particles; Use mixing and/or ultrasound; Fibres and needles form strong bridges
De-agglomeration requires higher energy input

Stabilisation

Agglomeration (increase in effective dp)

Do not promote too heavy collision of particles;
Use low particle concentration; Use de-flocculants (polyphosphates, organic polymers)

Table 2 Major parameters, effects and actions required.

Assessment of relevant measurement parameters

Solvents

The solvent should satisfy the following conditions:

  1. Be free of solid particles
  2. Having a good affinity for the particles
  3. Not dissolving the particles, or acting upon them by shrinking or swelling
  4. Not being reactive with the particles
  5. Having a different density from that of the sample in case of sedimentation

Usually deionised water is used for insoluble materials, although the pH of the deionised water could affect the zeta-potential of the particles. For materials which are insoluble but hardly wettable by water, adding an inorganic alcohol or a neutral detergent is effective, although detergents might cause foam formation. Organic solvents such as alcohols generally promote de-agglomeration of particles, whereas acetone or aromatics tend to promote particle adhesion on the cell wall, thus making measurements badly reproducible.

Dispersants

Dispersants have to be selected for each particle and solvent, but polyphosphates are usually applied for insoluble materials, with hexametaphosphate being most effective in most cases, although solutions cannot be stored for more than one day, since the dispersant loses its activity. The dispersant and its concentration should be selected on the basis of:

  1. The zeta-potential, preferably below - 60 mV
  2. Avoiding visual obscuration and sedimentation

Suspension aids

Of course, particle samples should be de-agglomerated before the measurement. The dispersant is hence important, but also ultrasonic and mechanical mixing are essential. The effect of sonication and mixing is further debated using experimental results.

The energy input during sonication should be limited to avoid particle disintegration. Also, care has to be taken to use an inert sonication tip, since cavitation can erode the tip, and since particles from the tip contaminate the sample suspension. The time of sonication could influence the measurement, with the size distribution shifting to fine sizes. Sonication should therefore be limited and additional mechanical stirring could be necessary to avoid re-agglomeration. If this nevertheless occurs, either a different dispersant or its higher concentration are needed.

Particle refractive index

The laser diffraction and scattering method has many advantages such as wide dynamic range, good reproducibility, easy operation and quick measurement. This is the reason why the laser diffraction and scattering method is most widely used. However, this method has a disadvantage which is difficult to handle, i.e., the input value of particle refractive index. When sample particles are of µm order or smaller, the value of the particle refractive index considerably affects the results as shown in Section 3. Data on the particle refractive index values are listed in handbooks such as “Handbook of Chemistry and Physics”.22 If the index is unknown, its measurement is very difficult.

Review of the relevant sample preparation conditions

Some previously mentioned and additional effects are summarized in Table 3.

Sample
Particle and solvent density, particle refractive index

Solvent
Type, refractive index, density and viscosity

Dispersant
Type and concentration

Sample suspension
Concentration, and temperature

Dispersion
Beaker size, dispersion device (Ultrasonication bath or tip), suspension volume, power, frequency, duration of Ultrasonication, tip material and size, tip position

Treatment of suspension
Duration from preparation to measurement,
dilution ratio of the suspension for measurement

Table 3 Parameters affecting the instrumental particle size measurement.

Results and discussion

Dispersants

Polyphosphate solutions are commonly used for insoluble powders, with hexameta-phosphate normally the most effective in many cases, although losing its activity in one day. As illustrated in the evolution of the zeta potential (ξ) with dosage of phosphates, the dispersing effect is active over a wide concentration range Figure 1. It is however clear that an over-dosage could lead to re-agglomeration rather than dispersion.

Figure 1 ξ -potential of Al2O3 versus dispersants and their concentration.

As alternative to polyphosphates, low-molecular weight organic dispersants are also widely used, including e.g. Daxad 11G,23 a condensation (Na+/K+) product of naphthalene sulphonic acids. Higher concentrations do not improve the dispersion and concentrations around 1% are recommended. Illustration of its use is given in Figure 2.

Figure 2 Effect of Daxad-concentration on measured particle sizes of SiO2.

Dispersion

Solvents: As stated in Section 2, the use of organic solvents can be necessary for water soluble powders. Alcohols, acetone or even aromatics can be used (as a function of the density of the powders). Figures 3-6 provide examples of organic solvents, with the poor performance of acetone probably due to particle adhesion on the cell wall.

Figure 3 Effect of solvents on the measured particle sizes of SiO2.

Figure 4 Effect of solvents on the measured particle sizes of saw dust.

Figure 5 Effect of solvents on the measured particle size of incinerator fly-ash.

Figure 6 Effect of solvents on the measured particle sizes of Sb2O3.

Amount of sample: As shown in Table 4, the amount of sample is critical, with excess sample leading to agglomeration.

Sample Amount (g)

d10 (µm)

d50 (µm)

d90 (µm)

0.5

(0.17)

0.25

0.73

1

(0.19)

0.30

0.80

2

(0.22)

0.36

0.85

4

(0.22)

0.43

0.90

Table 4 Effect of sample amount on the effect of agglomeration of TiO2

Mixing: Common laser diffraction equipment provides mechanical and ultrasonic mixing. The problems of non-appropriate use of the ultrasonic tip have already been described in Section 2. It is very important to maintain the amount of solvent in the bath at a required level, since mechanical and ultrasound mixing are affected by the liquid volume. The effect of sonication output power and duration are illustrated in Figure 7 & 8. Neither output power, nor duration of sonication have a significant effect within the applied range.

Figure 7 Influence of sonication output power (laser diffraction of Al2O3).

Figure 8 Influence of duration of sonication on the measured particle size.

Particle refractive index: A very important parameter when using laser diffraction is the particle refractive index. For samples of µm order or below, the refractive index influences the results considerably as depicted in Figure 9. It is therefore very important to make sure that the refractive index of the finer samples does not alter significantly.

Figure 9 Effect of the refractive index on the size distribution of Al2O3 (laser diffraction).

Presenting particle size distributions: The particle size distribution is usually presented as a histogram with the particle size on the linear or log-normal x-axis, and the fraction of each particle size range on the y-axis. This log-normal x-axis is used for the wide size-range, poly-disperse distributions. This y-axis fraction can be expresses either as a weight of particles (weight basis) or as a surface area of the particles (area basis) or as number of particles (number basis). Statistically defined particle size distributions are usually normalized and the total area of the histogram equals 1.

Each technique has its own basis where it yields the most reliable data. Most popular macroscopic methods (light scattering, acoustics) and fractionation methods (sedimentation, sieving, centrifugation) present data usually on a weight (same as volume) basis, on a logarithmic scale and as normalized distribution. If particle size distributions measured by different instruments, even by different methods, are expressed on the same basis and scale, then results are quite comparable, even for very broad particle size distributions.

Comparing common particle size analysers

Different particle analysers were compared for 2 powders, i.e. Al2O3 and SiC. The properties of both powders are given in Table 5.

 

SiC

Al2O3

Absolute density (kg/m³)

3130

3960

Specific surface area (m²/g)

20.5

1.9

Refractive index (-)

2.65

1.76

Dispersant and concentration (wt%)
ξ MathType@MTEF@5@5@+= feaagKart1ev2aaatCvAUfeBSjuyZL2yd9gzLbvyNv2CaerbuLwBLn hiov2DGi1BTfMBaeXatLxBI9gBaerbd9wDYLwzYbItLDharqqtubsr 4rNCHbGeaGqiVCI8FfYJH8YrFfeuY=Hhbbf9v8qqaqFr0xc9pk0xbb a9q8WqFfeaY=biLkVcLq=JHqpepeea0=as0Fb9pgeaYRXxe9vr0=vr 0=vqpWqaaeaabiGaciaacaqabeaadaqaaqaaaOqaaKqbakabe67a4b aa@383C@ -potential (mV)

Tri-sodium phosphate 0.025
-64

Sodium Hexametaphosphate 0.05
-97.5

Table 5 Sample properties and optimum dispersants.

The average values of 10 particle size measurements and the coefficients of variation are summarized in Table 6. As shown, the coefficients of variation of d50 measured by X-ray sedimentation, light obscuration and electrical sensing zone method are less than 10%. Photo-sedimentation has the worst relative accuracy. The coefficients of variation of d50 of the laser diffraction method are less than 20%.

Al203

X-Ray Sedimentation

Photo-Sedimentation

Light Obscuration

Electrical Sensing Zone

Malvern

d10 (µm)

(0.95)

(0.95)

1.16

1.16

0.71

CV (%)

2.80

14.20

5.80

8.30

35.9

d50 (µm)

1.81

1.69

2.88

2.16

2.10

CV (%)

3.00

12.60

7.20

4.80

12.70

d90 (µm)

3.68

4.13

4.89

4.07

4.69

CV (%)

5.20

41.80

3.20

4.60

9.60

SiC

X-Ray Sedimentation

Photo-Sedimentation

Light Obscuration

Electrical Sensing Zone

Malvern

d10 (µm)

(0.11)

(0.16)

0.63

(0.20)

(0.24)

CV (%)

(15.20)

(27.20)

3.50

(21.30)

34.50

d50 (µm)

0.47

0.47

1.02

0.68

0.64

CV (%)

21.70

39.40

6.90

10.40

18.00

d90 (µm)

1.92

1.60

3.12

2.71

1.96

CV (%)

10.80

34.70

17.70

14.50

31.20

Table 6 Comparison of results for Al203 and SiC.

Conclusion

The accuracy of the particle size analysis is largely dependent on a number of parameters, including the sample preparation, the degree of dispersion and the analysis itself. In most common suspension analysers, particles are dispersed in a solvent, mostly using a dispersant to lower the surface tension.

The use of organic solvents can be necessary for water soluble powders. Alcohols, acetone or even aromatics can be used (as function of the density of the powders). The results demonstrate that the type of solvent used affects the experimental results.

The amount of sample is critical, with excess sample leading to agglomeration. Common laser diffraction equipment provides mechanical and ultrasonic mixing. It is very important to maintain the amount of solvent in the bath at a required level, since mechanical and ultrasound mixing are affected. The effect of sonication output power and duration is negligible, within the applied test range.

A very important parameter when using laser diffraction is the particle refractive index, certainly for samples of µm order or below. It is therefore very important to make sure that the refractive index of the finer samples does not alter significantly.

If particle size distributions measured by different instruments, even by different methods, are expressed on the same basis and scale, then results are quite comparable, even for very wide particle size distributions.

The average values of 10 particle size measurements and the coefficients of variation demonstrate that the coefficients of variation of d50 measured are fair and mostly acceptable. Photo-sedimentation has the worst relative accuracy.

Acknowledgements

None.

Conflict of interest

The author declares no conflict of interest.

References

  1. Xi X, Cegla FB, Lowe M, et al. Study on the bubble transport mechanism in an acoustic standing wave field. Ultrasonics. 2011;51:1014–1025.
  2. Anand S, Nylk J, Dodds C, et al. Optical manipulation of aerosols using surface acoustic wave nebulisation. The International Society for Optical Engineering. 2011.
  3. Hasanzadeh H, Mokhtari–Dizaji M, Zahra Bathaie S, et al. Effect of local dual frequency sonication on drug distribution from polymeric nanomicelles. Ultrasonics Sonocheistry. 2011;18:1165–1171.
  4. Dukhin AS, Goetz PJ. Charaterisation of liquids, nano– and microparticulates, and porous bodies using ultrasound. Elsevier. 2010.
  5. Dukhin AS, Goetz PJ, Xiahua F, et al. Monitoring nanoparticles in the presence of larger particles in liquids using acoustics and electron microscopy. J Colloid Interface Sci. 2010;342(1):18–25.
  6. Naito M, Hayakawa O, Nakahira K, et al. Effect of particle shape on the particle size distribution measured with commercialequipment. Powder Technology. 1998;100(1):52–60.
  7. Saveyn H, Thu TL, Govoreanu R, et al. In–line comparison of particle sizing by static light schattering, time–of–transition and dynamic image analysis. Particle and Particle System Analysis. 2006;23(2):145–153.
  8. Tinke AP, Govoreanu R, Vanhoutte K, et al. Particulate system characterisation: Evaluation of particle size distribution data. American Pharmaceutical Review. 2007;10(5).
  9. Tinke AP, Govoreanu R, Vanhoutte K. Particle size and shape characterisation of nano and submicron liquid dispersion. American Pharmaceutical Review. 2006;9(5).
  10. Strokotov DI, Moskalensky AE, Nekrasov VM, et al. Polarised light–scattering profile–advanced characterization of non–spherical particles with scanning flow cytometry. Cytometry A. 2001;79(7):570–579.
  11. Araki M, Ebata A, Inoue T, et al. Sub–micron particle size measurements using multi–wavelength polarized lights. Hihon Kikai Gakkai Ronbunshu. Part B. 2010;76(770):1620–1628.
  12. Martos C, Coto B, Pena JL, et al. Effect of precipitation procedure and detection technique on particle size distribution of CaCO3. Journal of Crystal Growth. 2010;312(19):2756–2763.
  13. Shanthakumar S, Singh DN, Phadke RC. Methodology for determining particle size distribution characteristics of fly ashes. Journal of Materials in Civil Engineering. 2010;22(5):435–442.
  14. Celik H. Particle size analysis with X–ray sedimentation and laser light diffraction methods: General discussion of results obtained from two clay samples having different characteristics. Madencilik. 2009;48(4):3–18.
  15. Wan MH, Zhao MX, Ouyang JM. Research on particle size and size distribution of nanoparticles in urines by laser light scattering method. Spectroscopy and Spectral Analysis. 2009;29(1):217–221.
  16. Representation of results of particle size analysis, Part 1: Graphical representation. ISO; 9276-1:1998.
  17. Representation of results of particle size analysis, Part 5: Methods of calculation relating to particle size analyses using logarithmic normal probability distribution. ISO; 9276-5:2005.
  18. Particle size analysis, Dynamic light scattering (DLS). ISO; 22412:2008.
  19. Particle size analysis. Laser diffraction methods. 2009. ISO; 13320.
  20. Determination of particle size distribution by gravitational liquid sedimentation methods, Part 1: General principles and guidelines. ISO; 13317-1:2001.
  21. Particle size analysis. Image analysis methods, Part 1: Static image analysis methods. ISO; 20998.
  22. CRC. Handbook of Chemistry and Physics. Lide DR. 79th ed. 1998–1999.
  23. C.E.U. Speciality Chemicals.
Creative Commons Attribution License

©2017 Zhang, et al. This is an open access article distributed under the terms of the, which permits unrestricted use, distribution, and build upon your work non-commercially.