Case Report Volume 5 Issue 1
Cambridge University Hospitals NHS Foundation Trust, UK
Correspondence: GiotikasDimitrios, 30 Speedwell Close, CB1 9LF, Cambridge, Tel 448000000000
Received: April 04, 2016 | Published: May 12, 2016
Citation: Marko N, Dimitrios G, Matija K (2016) Massive Blood Loss during Graft Harvesting with the Reamer Irrigator-Aspirator. MOJ OrthopRheumatol 5(1): 00166. DOI: 10.15406/mojor.2016.05.00166
Reamer-Irrigator-Aspirator (RIA) is an FDA approved device for the reaming and preparation of the femoral medullary canal. It consists of a single-pass reamer that is connected to an irrigation bag and a suction device through two separated ports and tubes, allowing the surgeon to simultaneously irrigate and ream the canal and aspirate the reaming. The reaming can then be collected to a pot attached to the aspiration tube.
RIA was first introduced to reduce the incidence of fat embolism and thermal necrosis. Its indications have expanded since then, and it is now used for intramedullary debridement for the treatment of osteomyelitis and for bone graft harvesting.1,2
Several advantages of RIA, compared to graft harvesting from the iliac crest, have been reported: reduced postoperative pain, minimal donor site morbidity and larger volume of autologous bone graft.3,4 However, the RIA is not without complications: amongst others, the possibility of considerable blood loss has also been mentioned .1,2,4-6
We present two cases where massive bleeding was encountered during graft harvesting using RIA. Furthermore, we present our technical suggestions for the management of this potentially life threatening incident.
Case one
The first case was a 51-year-old male, otherwise healthy; who needed a bone graft as part of a second stage Masquelet technique for the treatment of acute traumatic bone loss from his proximal tibia. Three months earlier he had sustained a right open tibia fracture (Gustilo IIIB) with a large bone loss after a motorcycle road traffic accident. Despite the extensive bone and soft tissue loss, it was decided to salvage the limb. The initial debridement and spanning external fixator was followed by additional debridement of bone and soft tissues and internal fixation with locking plate. The missing extensor mechanism off patella was protected by tension by wire. Lastly, the tissue was covered with Latissimus Dorsi free muscle flap. For the graft harvesting, the femoral canal diameter was measured preoperatively on the CT and found to be 13mm. A 13.5 mm RIA head was used. During reaming, significant blood loss from the femoral canal was detected. The bleeding occurred within two minutes since the initiation of reaming. At that point, reaming and suction stopped and the RIA plastic tube was left in the canal as a tamponade. By calculating the difference between the fluid volume in the suction bag and the total administered volume of irrigating fluid we were able to see that approximately 2000 ml of blood had been lost. The patient developed mild tachycardia (less than 100/min) and was appropriately resuscitated by the anaesthetic team. Tranexamic acid was administered and the graft harvesting procedure was abandoned.
Case two
The second case was a 55-year-old male, otherwise healthy, who required bone graft as part of a second stage Masquelet technique for the treatment of persistent chronic osteomyelitis of the left distal femur. Osteomyelitis and the previous bone debridement had resulted in a cavitary bone defect on the anterolateral aspect of the bone. The contralateral femur was used for graft harvesting. No imaging of the femur was available preoperatively, so the femoral canal diameter was measured intraoperatively with the canal diameter measuring guide under image intensifier and found to be 13 mm. A 14 mm RIA head was used. The advancement of the RIA progressed slowly and with difficulty because of the hardness of the bone. After approximately six minutes, by calculating the difference between the fluid volume in the suction bag and the total administered volume of irrigating fluid, a blood loss of approximately 1500 ml was estimated. Reaming and suction were paused and the RIA plastic tube was left in the canal. The patient remained haemodynamically stable, tranexamic acid was administered and we carefully resumed the procedure 5 minutes later. We utilized standard femoral reamers to broaden the rest of the isthmus up to 13.5 mm and we then completed the reaming and graft harvesting with the RIA reamer. No further significant blood loss was noticed.
RIA has become increasingly popular as graft harvesting device. Although it is advantageous compared to standard iliac crest graft harvesting in terms of graft volume, donor side morbidity and possibly quality of graft,1,3,7 it is not without complications. Previous studies have highlighted the possibility of fractures and cortical breaching.1,2,4-6 Although the possibility of significant blood loss has also been mentioned in the majority of these studies,2,4-6,8 there are not many reported cases with massive blood loss. Quintero et al.6reported on 20 cases; one of them developed intraoperative hypovolemia. Based on their experience, they would expect a blood loss between 200 and 500 ml in RIA cases.The purpose of this study is to share our experience of this potentially disastrous event. In our cases the blood loss was significantly higher (1500 and 2000 ml) and happened very quickly. In our first case, it happened within the first two minutes of the reaming and before the first 3 l bag of irrigation fluid was finished. In the second case, it happened within six minutes of reaming.
These incidents occurred in otherwise healthy patients without history of bleeding disorders, coagulopathies or recent use of anti-DVT agents.
The technique that we used is the one described in the manufacturer’s Op-tech brochure and in the literature.4
Our experience confirms that massive bleeding is a complication that should be considered in any case scheduled for RIA. Also, appropriate precautions and measures should be implemented. Our technical recommendations based on this experience are:
The method of calculation of the total blood loss should be explained thoroughly.It is based on the deduction of the administered fluid from the total volume in the suction bag. It is easier to start the reaming with a new, full 3 lt irrigation bag and a completely empty suction bag. By deducting the volume given through the irrigation from the fluid volume in the suction bag we are able to calculate the blood volume lost at any given time. We stop and calculate the blood loss for every l of fluid administered through the irrigation. At the same time we take into account any blood and fluid on the floor or drapes.
Subsequently to these incidents we tested the use of cell saver in an attempt to re-administer the lost blood to the patient. It proved that the cell saver filter could not handle the large volume of fluid, as it is designed to be used with small amounts of fluids collected from the usual surgical field. It was necessary to pause the procedure several times to allow time for the cell saver to filter the fluid volume. This proved to be time-consuming and impractical.
We also tested the use of adrenaline solution 1:200.000 (we prefer 1 mg in 3 l of saline) diluted in the irrigating fluid. The use of adrenaline for haemostasis purposes has been successfully used for a long time in the shoulder arthroscopy in order to control intra-articular bleeding and improve visualization.9 Detailed information about its efficiency will be reported in our future adequately powered study. However, we think that, especially because it is a safety issue, it is worth mentioning here that the first data show reduced blood loss to less than 100 ml in the first four cases.
Based on our experience with the described cases we conclude that:
None.
None.
©2016 Marko, et al. This is an open access article distributed under the terms of the, which permits unrestricted use, distribution, and build upon your work non-commercially.