The application of micropulse transscleral contact-compression diode-laser cyclocoagulation for emergency reduction of intraocular pressure in patients with glaucoma

Kliuiev GO, Privalov AP, Luhova OM

Health Technology Ltd, Ukraine

Correspondence: Kliuiev GO, Health Technology Ltd, Ukraine

Received: March 27, 2018 | Published: April 6, 2018

Citation: Kliuiev G, Privalov AP, Luhova OM. The application of micropulse transscleral contact-compression diode-laser cyclocoagulation for emergency reduction of intraocular pressure in patients with glaucoma. MOJ Clin Med Case Rep. 2018;8(2):73-75. DOI: 10.15406/mojcr.2018.08.00244

Download PDF

Abstract

We analyzed the efficacy, safety and speed reduction of IOP which cannot be corrected by other methods by a single threshold (MTCCLCC) procedure in patients with different forms of glaucoma. After one session of MTCCDLC IOP decreased on average by 11.39 mmHg.

The relative decrease in IOP was 33.0% for the group of patients with POAG group and 27.6% for neovascular glaucoma. The overall success rate for all patients was 52%. The achieved efficacy and the absence of serious complications indicate that threshold MTCCLCC can be widely used in clinical practice as a means of treatment for a quick reduction in IOP.

Keywords: glaucoma, micropulse transscleral cyclocoagulation, laser, IOP

Introduction

Glaucoma was and remains one of the most relevant problems of ophthalmology. The percentage of blindness and low vision as a result of glaucoma remains stable and does not tend to decrease.1 Despite modern advances in drug therapy for glaucoma, in many cases, only laser and surgical treatment makes it possible to achieve a stable normalization of intraocular pressure and stabilize visual functions, and for so-called refractory glaucoma, laser surgery is generally a method of choice. In recent years, the arsenal of methods for the treatment of glaucoma has expanded due to the use of modern semiconductor lasers for transscleral cyclocoagulation.2,3

In the pathogenesis of glaucoma involved numerous factors - mechanical, vascular, metabolic. However, exactly the increase of IOP has a direct damaging effect on the structure of the optic nerve disc and leads, ultimately, to irreversible blindness. In severe cases, when the intraocular pressure cannot be reduced with maximum medication, and surgical treatment is associated with an increased risk due to possible complications, the use of laser transscleral cyclocoagulation is the method of choice.4 This method allows to reduce IOP quickly and effectively. However, the use of standard methods of laser exposure is fraught with significant complications, such as the development of hypotension, phthisis, hyphema, vitreous hemorrhage, uveitis, inflammation and other serious complications.5-7

Since 2006, we have begun to apply for these purposes micropulse transcleral contact-compression diode laser cyclocoagulation (MTCCDLC) with using a semiconductor laser at a wavelength of 810 nm. The MTCCDLC method allowed to increase safety and significantly reduce the level of complications.8 The increase in the number of patients requiring urgent reduction of ophthalmotonus and having a high risk of surgical intervention led us to study the possibility of using laser cyclocoagulation as an emergency procedure to reduce IOP. The aim of the research was to study the possibility of quickly, effectively and safely reducing IOP not corrected by other methods, a single procedure of MTCCDLC in patients with different forms of glaucoma.

Materials and methods of research

The research was conducted in 109 patients (115 eyes), the average age of which is 65.33±13.75. 52 of them were women and 57 of them were men. All patients received the maximum medication. MTCCDLC was used in cases when drug therapy did not lead to normalization of IOP or there were contraindications to surgical treatment. There was used the coagulator developed by us for transcleral cyclocoagulation problems with a semiconductor laser with a wavelength of 810 nm. The coagulator is adapted with a quartz-polymer fiber with a diameter of 600 μm. The design of the transscleral probe provided a smooth change in the force of 0.1-0.5 N with the adjustment screw of the spring device. The radiation power was limited to 1.0-1.4 W, duration 1 s, the number of applications 30-40 along the circumference 270* in 1, 8-2 mm from the limb. The axis of the transscleral probe was oriented perpendicular to the scleral surface and deviated from the optic axis of the eye. Initial increase in fiber pressure was established by 0.1 N and increased for elderly patients. Cyclocoagulation was performed under combined epibulbar anesthesia with 0.5% alkane (3 fold installation) and parabulbar anesthesia with marcaine(5 mg / ml)-3ml.

The results obtained and their discussion

Using the Mann-Whitney test, IOP was compared before surgery with IOP of the remaining time intervals. M±m –average value with standard error, N–number of eyes, Min/Max–minimum/maximum value of IOP. The Mann-Whitney test showed that statistically significant (significance level p<0.05) is IOP a day after the operation and IOP at discharge, i.e. there is a difference between IOP in patients diagnosed with NVG and POAG. That means, that in the group of patients with NVG and POAG, a significant difference is present between the initial IOP, the pressure measured one day after the operation and IOP at the end of the time. If to pay attention to the indicators of mean values, then after one session of MTCCDLC, IOP decreased on average 11.39 mmHg (Figure 1) (Table 1).

Figure 1 1–IOP before operation, 2–IOP after one day of operation, 3 – IOP after some days of operation, 4 – IOP discharge.

 

Average value
M±m

N

Min

Max

Standard deviation

Significance level р*

IOP before

32,03±0,58

115

25

58

6,61

 

IOP after one day

23,62±0,53

108

15

38

5,74

0,013368

IOP intermediate

22,18±0,58

91

15

38

5,83

0,072278

IOP discharge

20,64±0,36

115

15

36

4,06

0,016623

Energy

48,75±2,93

111

20,00

116,00

23,37

 

Table 1 Change in IOP in patients with diagnoses POAG and NVG after MTCCDLC

The relative decrease in IOP was 33.0% for the POAG group and 27.6% for the neovascular glaucoma. The overall success rate for all patients was 52%. At an average energy in the pulse of 1.3 J, the average summary radiation energy per session was 39.6 J, which is lower than the energies used by standard methods. The manifestation of complications that are caused by explosive effects arising from the use of elevated capacities in the implementation of standard cyclocoagulation techniques have been virtually eliminated, since the power and average energy in the pulse for MTCCDLC are much lower than necessary for the manifestation of these effects.

Conclusion

Micropulse transcleral contact-compression diode laser cyclocoagulation (MTCCDLC) with the achieved efficacy and the absence of severe complications can be widely used in clinical practice as an emergency aid for the rapid reduction of IOP in patients with different forms of glaucoma with high IOP, not corrected by medical therapy.

Acknowledgements

None.

Conflict of interest

The author declares there is no conflict of interest.

References

  1. Quigley HA, Broman AT. The number of people with glaucoma worldwide in 2010 and 2020. Br J Ophthalmol. 2006;90(3):262–267.
  2. Bloom PA. Ciliary body laser treatment for glaucoma: a contemporary view. Glaucoma world. 2000;5(20):10–17.
  3. Schlote T, Derse M, Rassman K. Efficacy and safety of contact transscleral diod laser cyclophotocoagulation for advanced glaucoma. J glaucoma. 2001;10(4):294–301.
  4. Gaasterland DE, Pollack IP, Gaasterland DE. Initial experience with a new method of laser transscleral cyclophotocoagulation for ciliary ablation in severe glaucoma. Trans Am Ophthalmol Soc. 1992;90:225–46.
  5. Volkov VV. Transscleral diodmicrolaser cyclophobic coagulation in the treatment of complicated forms of glaucoma. Volkov VV, Kachanov AB, 6th congress of ophthalmologists in Russia: abstracts. M. 1994. 215 p.
  6. Kachanov AB. Diode-laser transscleral contact cyclocoagulation in the treatment of various forms of glaucoma and ophthalmic hypertension: author's abstract. diss. for the competition uch. degree of Cand. Honey. Science. 995:21.
  7. Egbert PR, Fiadoyor S, Budenz DL. Diod transscleral cyclophotocoagulation as a primary surgical treatment for primary open-angle glaucoma. Arch Ophthalmol. 2001;119(3):345–350.
  8. Klyuyev GO. Compression transscleral probes for laser cyclocoagulation.Ophthalmol. Journal. 2007;6:62–65.