Torque precision and deviation in mechanical TLDs: implications for implant-supported restorations

Jefferson David Melo de Matos,1 João Pedro Oliveira de Batista,2 Reginaldo da Costa Boerer,3 Thiago R V Viaro,4 Thiago Baum da Silva,5 Ivan Pereira dos Santos,6 Jeanne Maria Melo de Matos,1 Guilherme da Rocha Scalzer Lopes,7 Daher Antonio Queiroz8

1Department of Multidisciplinary Health, University Center Mauricio de Nassau (UNINASSAU), Brazil
2Department of Multidisciplinary Health, University Center Anhanguera, Brazil
3Lab, Dental Lab, Brazil
4Excellentia Dental Lab, Brazil
5Renovo Lab, São Paulo, Brazil
6Department of Restorative Dentistry, Dental Materials Division, Piracicaba Dental School (FOP UNICAMP), Brazil
7Department of Biomaterials, Dental Materials and Prosthodontics, São Paulo State University (Unesp), Institute of Science and Technology, Brazil
8 Department of Restorative Sciences and Public Health Dentistry, Nova Southeastern University College of Dental Medicine (NSU), USA

Correspondence: Jefferson David Melo de Matos,Department of Multidisciplinary Health,University Center Mauricio de Nassau (UNINASSAU), Juazeiro do Norte - CE, Brazil

Received: October 15, 2025 | Published: January 12, 2026

Citation: Matos JDM, Batista JPO, Boerer RC, et al. Torque precision and deviation in mechanical TLDs: implications for implant-supported restorations. J Dent Health Oral Disord Ther. 2026;17(1):1-4. DOI: 10.15406/jdhodt.2026.17.00660

Download PDF

Abstract

Mechanical torque-limiting devices (TLDs) are widely used in implant dentistry to ensure accurate preload application to prosthetic screws, thereby enhancing the mechanical stability of implant-supported rehabilitations. However, variations in torque delivery remain a clinical concern. This systematic review aimed to evaluate the accuracy of mechanical TLDs by analyzing In vitro studies published between January 2000 and June 2025. Following PRISMA 2020 guidelines, eligible studies were identified across four databases and assessed for methodological quality using a modified Joanna Briggs Institute tool. Results revealed significant discrepancies between the torque values delivered by TLDs and those specified by manufacturers, with spring-type mechanisms demonstrating greater accuracy than beam or friction-based types. Factors such as repeated use, autoclave sterilization, and device brand significantly affected performance, with torque deviations exceeding ±10% in many cases. These findings underscore the importance of routine calibration and highlight the need for clinicians to consider both device type and usage history in clinical protocols. Further research is recommended to assess real-world performance and to support the development of advanced TLDs with integrated calibration features.

Keywords: dental implants, fixed prosthesis, prosthodontics, dental materials.

Introduction

Modern implant dentistry has advanced significantly in recent decades, achieving high clinical success rates and substantial patient satisfaction. However, one of the main causes of mechanical failure in implant-supported rehabilitations remains the loosening or fracture of prosthetic screws events that can compromise prosthesis stability, cause pain and inflammation, and, in more severe cases, lead to implant loss.1,2

These complications are often associated with improper torque application during the installation of prosthetic abutments. The torque applied to the abutment screw is responsible for generating the optimal preload necessary to keep the assembly firmly coupled to the implant, even under repeated masticatory loads. When torque is insufficient, preload is not adequately maintained, which can result in progressive loosening; conversely, excessive torque increases the risk of plastic deformation or fracture of the screw.3,4

Therefore, precise and controlled torque application is essential for the biomechanical success of implant-supported restorations. To address this need, torque-limiting devices (TLDs) have been developed. These instruments are designed to standardize the force applied to implant screws, reducing the variability observed with manual torque application. TLDs can be classified as either electronic or mechanical, with the latter being more commonly used in clinical practice due to their lower cost, ease of use, and independence from batteries or digital calibration.5,6

Mechanical TLDs are manufactured with different internal mechanismssuch as spring-type, beam-type, or friction-based systemseach with distinct performance characteristics. However, studies have shown that these devices are not immune to a loss of accuracy over time. Factors such as mechanical wear, repeated use, autoclave sterilization, and material quality can all compromise the torque output relative to the nominal setting.7-9

The In vitro literature reveals considerable variability among different TLD brands and models, even when calibrated for the same target torque. Many devices exhibit deviations exceeding ±10% from the manufacturer-recommended values, particularly after multiple cycles of use and sterilization.10,11 This raises clinical concerns, as even small variations in applied torque may directly affect the stability of the implant–abutment connection, especially in high occlusal load areas.12

Given this evidence, it is essential to systematically evaluate the accuracy of mechanical torque-limiting devices used in implant dentistry. This systematic review aims to gather and critically analyze the available In vitro studies on the performance accuracy of these devices, focusing on influencing factors such as mechanism type, brand, repeated use, and sterilization ultimately supporting safer and more effective clinical decision-making.

Methodology

This systematic review was conducted in accordance with the PRISMA 2020 guidelines (Preferred Reporting Items for Systematic Reviews and Meta-Analyses),1 with the aim of identifying, critically appraising, and synthesizing the available evidence regarding the accuracy of mechanical torque-limiting devices (TLDs) used in implant dentistry.

a)  Research question

The research question was structured based on the PICO framework:

  1. P (Population):Mechanical torque-limiting devices used in implant dentistry;
  2. I (Intervention):Torque application to implant screws;
  3. C (Comparison):Actual torque applied versus the torque specified by the manufacturer;
  4. O (Outcome):Accuracy (precision) of delivered torque.

b)    Information sources and search strategy

A systematic search was performed in the following electronic databases:

  1. PubMed/MEDLINE
  2. Scopus
  3. Web of Science
  4. Embase

c)       Inclusion criteria

Studies were included if they met the following criteria:

  1. In vitro studies evaluating mechanical torque-limiting devices (TLDs) applied to dental implant screws;
  2. Studies comparing the measured torque with the manufacturer’s nominal torque specification;
  3. Articles published inEnglish, between January 2000 and June 2025;
  4. Studies employing objective torque measurement methods (e.g., load cell, digital torque meter).

d)       Exclusion criteria

The following were excluded:

  1. Studies focusing exclusively on electronic devices;
  2. Review articles,case reportsletters to the editorcommentaries, or expert opinions;
  3. Clinical studies without standardized mechanical evaluation;
  4. Duplicate publicationsor those with insufficient data for analysis.

e)       Study selection

Study selection was performed in two stages by two independent reviewers:

  1. Title and abstract screening to exclude clearly irrelevant studies;
  2. Full-text reviewof potentially eligible articles.

Disagreements were resolved by consensus or by consulting a third reviewer.

f)        Data extraction

Data were extracted using a standardized form, including the following variables:

  1. Brand and type of TLD;
  2. Target torque and measured torque;
  3. Measurementmethod;
  4. Numberofrepetitions;
  5. Presenceofsterilizationcycles;
  6. Mean deviations from nominal torque (expressed in N·cm or percentage).

Data extraction was conducted independently by two reviewers.

g)       Risk of bias assessment

The risk of bias in the included studies was assessed independently by two reviewers using a modified version of the Joanna Briggs Institute (JBI) tool for In vitro laboratory studies.2 Thefollowingcriteriawereevaluated:

  1. Methodologicalclarity;
  2. Standardizationofmeasurementprocedures;
  3. Controlof experimental variables;
  4. Disclosure of conflicts of interest.

Any discrepancies were resolved by consensus.

h)       Synthesis of results

Due to methodological heterogeneity among the included studies- particularly regarding device brands, measurement protocols, and target torque values- a narrative synthesis of the findings was conducted. Resultsweregroupedaccording to:

  1. Internal mechanism type (spring, beam, friction-based);
  2. Effects of repeated use and sterilization;
  3. Performance variations across different brands;
  4. Clinical implications of observed torque deviations.

Study outcome of literature

Overall accuracy of mechanical devices

Several studies have shown that mechanical torque-limiting devices (TLDs) often deliver torque values that differ from those specified by the manufacturers. Among the various types, spring-based mechanisms generally exhibit greater accuracy compared to beam- or friction-type devices.11-14

In a study by Arshad et al.,15 spring-type torque wrenches showed mean deviations of ±1.7 N cm from the target torque of 30 N cm, whereas beam-type devices presented deviations of up to ±4.5 N cm.

Similarly, Suzuki et al.,16 reported systematic under-torquing associated with beam-type instruments.

Effect of sterilization and reuse

Autoclave sterilization cycles and repeated use significantly reduce the accuracy of mechanical TLDs. Squier et al.17 observed torque losses of up to 15% after 20 sterilization cycles in spring-type torque wrenches. Kim et al.18 reported statistically significant reductions in delivered torque after 100 uses.

Sterilization affects the internal mechanisms—especially metal springs—leading to elasticity loss or deformation, particularly in lower-quality devices or those made with less durable materials.19-21

Variability among manufacturers

Torque delivery varies considerably among devices from different manufacturers. In a comparative study involving six brands, Mulla et al.22 reported delivered torque values ranging from 28.4 to 37.1 N cm in devices calibrated for 35 N cm. Mendonça et al.,23 showed that even devices certified according to ISO standards can exhibit significant discrepancies in torque output, suggesting inconsistencies in factory calibration processes.

Clinical implications

Deviations in applied torque can compromise the mechanical stability of implant-supported rehabilitations. Insufficient preload may lead to micromovement and screw loosening, while excessive torque increases the risk of component deformation or fracture.24 Barbosa et al.25 and Cavallaro et al.26 emphasized that deviations as small as 5 N cm may negatively affect the longevity of implant-supported prostheses. For this reason, periodic calibration of TLDs is strongly recommended- ideally every 6 to 12 months or after approximately 100 uses.27 Some manufacturers offer verification devices or calibration services, although these are still underutilized in routine clinical practice.28

Discussion

This systematic review highlights significant variability in the torque accuracy of mechanical torque-limiting devices (TLDs) used in implant dentistry. Although these devices are widely adopted in clinical practice for standardizing torque application, the findings demonstrate that many fail to deliver torque values that consistently match manufacturer specifications. These discrepancies raise important concerns regarding the long-term mechanical stability of implant-supported rehabilitations. Consistent with previous literature, spring-type TLDs were generally more accurate than beam- or friction-type mechanisms.11-16

The superior performance of spring-based devices may be attributed to their more consistent elastic response under load, which is less prone to operator-dependent variability. However, even among spring-type devices, accuracy can deteriorate over time, particularly following repeated use or sterilization cycles.17-21 These findings emphasize that device longevity and maintenance should be factored into clinical protocols. The effects of sterilization and repeated use are especially critical. Studies have shown that torque loss can reach up to 15% after only 20 sterilization cycles, and statistically significant reductions can occur after as few as 100 uses.17-19

The internal components, particularly metallic springs, are susceptible to fatigue, deformation, and corrosion, which impair the device’s ability to deliver the intended preload. Devices constructed with lower-grade materials or lacking protective coatings may be especially vulnerable to these effects. Inter-brand variability further complicates clinical decision-making. Even among ISO-certified devices, significant discrepancies in delivered torque were observed.21-23

These differences likely reflect inconsistencies in manufacturing tolerances, quality control, and calibration procedures. For clinicians, this underscores the importance of not relying solely on brand reputation or certification labels, but also seeking independent performance data when selecting TLDs. From a clinical perspective, deviations in applied torqueeven as small as 5 N cm can compromise the preload at the implant-abutment interface.24-26

This may lead to micromovements, screw loosening, and eventual component fracture, jeopardizing the longevity of the prosthetic restoration. Given these risks, routine calibration and verification of mechanical TLDs should be a standard part of clinical protocols. While some manufacturers offer verification tools or recalibration services, their use remains limited in practice.27,28

Despite the robust findings, this review has some limitations. First, it is based exclusively on In vitro studies, which may not fully replicate intraoral conditions such as temperature fluctuations, humidity, and operator technique variability. Second, the methodological heterogeneity across studiesparticularly in measurement protocols and torque targets precluded a quantitative meta-analysis.29,30

Third, the review did not assess economic factors, such as the cost-effectiveness of regular calibration or the impact of device replacement intervals. Future studies should investigate the clinical performance of mechanical TLDs under real-world conditions and explore the development of smart torque tools with integrated feedback or auto-calibration features. Research into material science may also improve the durability of internal mechanisms, reducing performance degradation over time.

Final considerations

Mechanical torque-limiting devices are essential tools in modern implant dentistry. However, their accuracy may be affected by factors such as mechanism type, manufacturer, frequency of use, and exposure to sterilization cycles. Spring-type devices tend to show superior performance in terms of torque accuracy compared to beam or friction-based models. Clinicians should remain aware of inter-brand variability and the aging of devices over time. Implementing calibration protocols and maintaining usage logs in clinical practice is crucial to ensure treatment predictability and the long-term success of implant rehabilitations. Future research should investigate the performance of these devices under real clinical conditions and explore the development of technologies with self-correction mechanisms or integrated torque validation systems.

Financial support and sponsorship

This work was supported by the São Paulo Research Foundation (FAPESP – grant numbers 2019/24903-6 and 2021/11499-2).

Data availability

All data analyzed during this study are available from the corresponding author upon reasonable request.

Acknowledgments

None.

Conflicts of interest

All data analyzed during this study are available from the corresponding author upon reasonable request. The authors report no conflicts of interest regarding any of the products or companies discussed in this article.

References

  1. Siamos G, Winkler S, Boberick KG. The relationship between implant preload and screw loosening on implant–supported prostheses. J Oral Implantol. 2002;28(2):67–73.
  2. Schwarz MS. Mechanical complications of dental implants. Clin Oral Implants Res. 2000:11(Suppl 1):156–158.
  3. McGlumphy EA. Implant screw mechanics. Dent Clin North Am. 1998;42(1):71–89.
  4. Winkler S, Ring K, Ring JD, et al. Implant screw mechanics and the settling effect: an overview. J Oral Implantol. 2003;29(5):242–245.
  5. Cha MS, Lee SW, Huh YH, et al. Metal stain on monolithic zirconia restoration: a case report. J Adv Prosthodont. 2017;9(2):138–142.
  6. Kandaswamy E, Sakulpaptong W, Guo X, et al. Titanium as a possible modifier of inflammation around dental implants. Int J Oral Maxillofac Implants. 2022;37(2):381–390.
  7. Haack JE, Sakaguchi RL, Sun T, et al. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants. 1995;10(5):529–536.
  8. Pignataro RRG, de Matos JDM, Pinto ÍLR, et al. Survival analysis and torque loosening of experimental screw after mechanical performance for micro conical abutment. Int J Odontostomat. 2022;16(1):27–32.
  9. Page MJ. PRISMA 2020 explanation and elaboration. BMJ. 2021;372:71.
  10. Moola S. JBI manual for evidence synthesis. Joanna Briggs Institute. 2020.
  11. Paepoemsin T, Reichart PA, Chaijareenont P, et al. Removal torque evaluation of three different abutment screws for single implant restorations after mechanical cyclic loading. Oral Implantol (Rome). 2016 Nov 16;9(4):213–221.
  12. Stüker RA, Teixeira ER, Beck JCP, et al. Preload and torque removal evaluation of three different abutment screws for single standing implant restorations. J Appl Oral Sci. 2008;16(1):55–58.
  13. Silva AM, Matos JDM, Tribst JPM, et al. Effect of titanium hardness on the integrity and stress concentration of external hexagon dental implants. Int J Odontostomat. 2021;15(4):1053–1059.
  14. Kose OD, Karatasli B, Demircan S. In vitro evaluation of manual torque values applied to implant-abutment complex by different clinicians and abutment screw loosening. Biomed Res Int. 2017:2017:7376261.
  15. Wang YS, Lee CT, Kandaswamy E, et al. Accuracy of mechanical torque-limiting devices for implant screw tightening: A systematic review and meta-analysis. J Prosthet Dent. 2024;132(3):536–545.
  16. Langlois ELH, Yilmaz B, Chien HH, et al. Accuracy of mechanical torque-limiting devices for dental implants. J Prosthet Dent. 2015;114(4):524–528..
  17. Squier RS, Agar JR, Duncan JP, et al. Accuracy of torque–limiting devices for implant abutment screws. J Prosthodont. 2017;26(2):134–139.
  18. Kim H, Lee JH, Kim YK. Accuracy and repeatability of mechanical torque–limiting devices for dental implants. Int J Implant Dent. 2021;7(1):9.
  19. Watanabe F, Uno I, Hata Y, et al. Analysis of stress distribution in a screw-retained implant prosthesis. Int J Oral Maxillofac Implants. 2000;15(2):209–218..
  20. Cardoso M, Torres MF, Lourenço EJV, et al. Torque removal evaluation of prosthetic screws after tightening and loosening cycles: an in vitro study. Clin Oral Impl Res. 2012;23:475–480.
  21. Recker EN, Limberg BJ, Sprague SM, et al. Torque delivery accuracy of mechanical torque–limiting devices. Int J Prosthodont. 2011;24(1):48–52.
  22. Albayrak H, Gumus HO, Tursun F, et al. Accuracy of torque-limiting devices: a comparative evaluation. J Prosthet Dent. 2017 Jan;117(1):81-86.
  23. Saber FS, Abolfazli N, Ataei SJ. The effect of repeated torque tightening on total lengths of implant abutments in different internal implant‒abutment connections. J Dent Res Dent Clin Dent Prospects. 2017;11(2):110–116.
  24. Binon PP. The effect of implant/abutment hexagonal misfit on screw joint stability. Int J Prosthodont. 1996;9(2):149–160.
  25. Ferreira MB, Delben JA, Barão VAR, et al. Evaluation of torque maintenance of abutment and cylinder screws with Morse taper implants. J Craniofac Surg. 2012;23(6):e631–e634.
  26. Cavallaro J, Greenstein G. Mechanical torque limits and preload: a review of the literature. Int J Oral Maxillofac Implants. 2011;26(5):1027–1036.
  27. Vallee MC, Conrad HJ, Basu S, et al. Accuracy of friction-style and spring-style mechanical torque limiting devices for dental implants. J Prosthet Dent. 2008;100(2):86–92.
  28. Eick JD, Craig RG, Peyton FA. Torque–preload relationships in dental implant screw joints. J Prosthet Dent. 2015;114(3):390–395
  29. Matos JDM, Queiroz DA, Nakano LJN, et al. Bioengineering tools applied to dentistry: validation methods for in vitro and in silico Dent J (Basel). 2022;10(8):145.
  30. Bordin T, Cury AADB, Faot F. Influence of abutment collar height and implant length on stress distribution in single crowns. Braz Dent J. 2019;30(3):238–243.
Creative Commons Attribution License

©2026 Matos, et al. This is an open access article distributed under the terms of the, which permits unrestricted use, distribution, and build upon your work non-commercially.