Submit manuscript...
Journal of
eISSN: 2373-4396

Cardiology & Current Research

Correspondence:

Received: January 01, 1970 | Published: ,

Citation: DOI:

Download PDF

Abstract

SelectSecure 3830 pacemaker leads have been shown to be safer and more advantageous in the short term for pediatric pacemaker patients compared to previous pacemaker lead designs. The two case studies outlined below illustrate potential pitfalls in the medium to long term care of patients with this pacing lead. Further research on the long term risks/benefits of these leads must be performed to determine the long term functionality of the SelectSecure 3830 lead.
Keywords: Pediatrics; Pacing leads; SelectSecure 3830; Lead advancement

Introduction

The SelectSecure3830 steroid eluting, bipolar, non-retractable screw-in transvenous lead (Medtronic, Inc.) provides a thinner diameter lead for pacemaker patients due to the lack of a central stylet lumen. This 4.1 F lead has been shown in preliminary data to be both safe and efficacious [1] without increasing the risk of venous thrombosis [2]. Long term safety results for the lead are still pending. Pediatric pacemaker patients present a unique challenge to lead placement as they have significant linear growth often requiring leads to be advanced or exchanged. Lead extraction/advancement has the chance for major complications particularly with older leads [3].

Case Report

We report our institutional experience with lead advancement with the SelectSecure 3830 lead in two pediatric patients. In the first case, a 5-year-old female with a history of complete heart block underwent implantation of a single chamber transvenous pacemaker with a Medtronic 3830-59 SelectSecure ventricular pacing lead at the age of 3. She was brought back to the EP lab for advancement of her 2 year old ventricular lead due to growth related ventricular lead tension. Under fluoroscopic guidance, the lead was easily advanced into the vasculature gaining several inches. During that same procedure, the patient was upgraded to a dual chamber system with a second Medtronic 3830-59cm lead placed in the atrium.
The second case was a 12-year-old female with history of complete heart block who underwent placement of a dual chamber transvenous pacemaker at 8 years of age with two Medtronic 3830-59cm pacing leads. She returned to the EP laboratory with 4 year old leads due to growth related tension of the atrial lead. Under fluoroscopic guidance, the atrial lead was advanced further into the right atrial appendage with some difficulty. The following morning, her atrial lead was found to be non-functional but intact. Attempts to either advance or extract the lead with simple traction were unsuccessful, and as such a new 3830-59cm lead was implanted in the right atrial appendage leaving sufficient slack for growth. The ventricular lead was unable to be advanced due to scar tissue at the innominate vein/SVC junction. Follow-up at 1 month showed stable atrial pacing/sensing thresholds.

Discussion

Previous research on the Select Secure 3830 lead has been focused on safety during implantation and short/medium term follow up. The results of our case studies show that through the medium and long term this lead will be subject to the usual problems of transvenous leads and may not be easily advanced as hoped. The lead inconsistencies started to appear with longer term follow up, at 4 years in one of our cases. A definitive conclusion on the risks and benefits of lead advancement/extraction with the SelectSecure 3830 lead cannot be concluded from the reported cases. More studies are needed to evaluate the long term functionality of the SelectSecure 3830 stylet-less lead.

References

  1. Gammage MD, Lieberman RA, Yee R, Manolis AS, Compton SJ, et al. (2006) Multi-center clinical experience with a lumenless, catheter-delivered, bipolar, permanent pacemaker lead: implant safety and electrical performance. Pacing Clin Electrophysiol 29(8): 858-865.
  2. Gabbarini F, Agnoletti G (2010) Selective-site pacing in paediatric patients: use of the SelectSecure System and risk of vein occlusion. Europace 12(9): 1286-1289.
  3. Cecchin F, Atallah J, Walsh EP, Triedman JK, Alexander ME, et al. (2010) Lead extraction in pediatric and congenital heart disease patients. Circ Arrhythm Electrophysiol 3(5): 437-444.
Creative Commons Attribution License

© . This is an open access article distributed under the terms of the, which permits unrestricted use, distribution, and build upon your work non-commercially.