Mini Review Volume 3 Issue 2
University Claude Bernard Lyon1, France
Correspondence: Latifa Fakri Bouchet, University of Lyon, University Claude Bernard Lyon1, UMR CNRS 5280, Institute Des Sciences Analytiques (ISA), 5 Rue De La Doua, Villeurbanne, France, Tel 06031842176
Received: July 31, 2017 | Published: September 15, 2017
Citation: Fakri-Bouchet L, Zahraoui M. Implantable microcoil for in-vivo magnetic resonance spectroscopy. Int J Biosen Bioelectron. 2017;3(2):237–240. DOI: 10.15406/ijbsbe.2017.03.00056
Nuclear Magnetic Resonance (NMR) allows a non-destructive study of molecular structures by the measurement of their atomic nuclei magnetization. Nonetheless, NMR technique still suffers from the sensitivity weakness compared to all other spectroscopy approaches. Therefore, the optimization of detection Radio Frequency (RF) coils and the reduction of the thermal noise will lead to an increase in signal resolution particularly around the regions of interest (ROI) within the living sample. In the case of sample with limited mass and volume, the sensitivity may be improved for when the size of the RF coil matches the sample size. In this study, we developed a microcoil (500x1000μm) dedicated to the NMR micro imaging (MRI) and micro localized spectroscopy (MRS) for biomedical applications on small rodent models of human diseases. Our NMR microcoil offer a sensitivity increased by 2.5, and limits of detection (LOD) of the main cerebral metabolites estimated by MRS in-vitro measurements in good concordance with analytical values given in literature. The realization of compact sensing system based on microcoil will open the door to numerous medical and industrial.
Keywords: nuclear magnetic resonance, spectroscopy, in-vivo, region of interest, limit of detection, radiofrequency, microcoil
NMR, nuclear magnetic resonance; MRS, micro localized spectroscopy; MRI, micro imaging; ROI, regions of interest; LOD, limits of detection; RF, radio frequency; µcoil, microcoil
The Nuclear Magnetic Resonance Spectroscopy (MRS) is the study metabolic processes in different biological and chemical samples as tissues, cell cultures, and protein structures. MRS fulfils an important role through its ability to deduce structural information and to withdraw microscopic data from the intermolecular dynamics. The NMR Approach has expand its capacities to cover furthermost medical and chemical diagnosis applications. However, NMR sensitivity still lags significantly behind most other analytical techniques by a factor of 100-1000, mainly for mass-limited and concentration-limited samples.1 The observation of metabolic processes using short spin-spin relaxation decay becomes possible via short echo-time localization pulse-sequences. Yet, spectra quantification is obstructed by overlapping metabolite resonances and by low Signal-to-Noise Ratio (SNR) due to the limited size of the observed volume. Hence, the analysis of sample volumes of nanoliter order point toward the improvement of Radio-Frequency (RF) coils having dimensions of several micrometers.2,3 The sensitivity of an NMR receiver coil is expressed as a ratio of the SNR and the concentration of the measured substance:4
(1)
Where Ci is the concentration of the measured substance in mol/kg. The SNR is the fundamental quantifier of microcoil’s sensitivity due to the existence of the Noise as a limiting factor. SNR is an empirical relation transcribed as:
(2)
The SNR could also be depicted analytically by the physical parameters of the microcoil and the surrounding setting:
(3)
where kB is the constant of Boltzmann, TS is the ambient temperature, Rs is the coil conductors equivalent resistance, and is the receiver bandwidth. Two main components could be differentiated: (1) The environment components are the static magnetic field B0, the Larmor precession frequency ω0 proportional to B0 and the sample volume Vs. (2) The geometrical components are B1/i which is equivalent to the NMR signal by the principle of reciprocity and symbolizes the thermal noise produced by the coil in a given bandwidth.5
The experimental equipment providing the high static field B0 had are stirring toward 20T magnetic resonance intended for studying human brain still they rarely available for typical research.6 Thus, the geometrical parameters are optimized to attain a maximum B1 and to reduce the coil dimensions (minimum Rs) so a low noise Signal could be detected.
Microcoil geometry effect on sensitivity
The design of a planar sub-milimetric coil recommend defining firmly the dimension so the detection size of the Radio Frequency coil matches the sample size. Spiral configurations, as shown in Figure 1, could be the finest solution for in-vivo implant examinations.7 From the optimization results, the square coil (1000x1000µm2) have displayed a high sensitivity for the observation of tiny sample of 1000µl. Although, the race track shape have numerous physical advantage such as the round corners, which reduce the effect of hotspots that can burn the sample, and especially the half surface area compared to the square coil, the thing that will reduce lesion size after the implantation inside the brain. Table 1 represent a complete comparison between geometric shapes of the microcoil considering the ratio called quality factor Q of the receiving coil (Figure 2):
(4)
Where the equivalent resistance and the admittance Lω, and RAC is the the µcoil Alternative Current resistance. All Magnetic Resonance Imaging (MRI) and Magnetic Resonance Spectroscopy (MRS) experiments were performed using a 4.7T-Bruker Biospec System, the resonance frequency is about f0 = 200 MHz.
Configuration |
ns (Spires) |
W (µm) |
∆ (µm) |
RAC (Ω) |
L (nH) |
Q |
B1max |
Square |
6 |
29 |
20 |
0.98 |
33.2 |
42.5 |
1.53 |
Circular |
6 |
29 |
20 |
0.82 |
28.2 |
43.3 |
1.31 |
Rectangular |
4 |
22 |
20 |
0.64 |
12.7 |
25.1 |
0.62 |
Ellipsoidal |
4 |
22 |
20 |
0.59 |
12.3 |
26.3 |
0.59 |
Table 1 Configurations study of the different geometries with the optimal parameters
In-Vitro study & the comparison with commercial surface coil
The microcoil performance was compared to a Rapid-Biomed surface-coil (3.5cm diameter) designed for 1H of rat brain. For proton metabolites detection, we have evaluated their sensitivities and limits of detection (LOD) as defined in4 and determined the factor of gain (FOG) of the microcoil approach defined as
(5)
MRS measurements were set on a phantom containing ten cerebral metabolites solutions. The chosen planar homemade micro-coil had ellipsoidal geometry (1000x500µm²), had four concentric turns, (22µm trace width, 46µm trace thickness and 20µm spacing between turns, quality coefficient=24). The planar micro-coil was immersed in the phantom. Both coils were used for signal detection only and RF transmission was accomplished using a Rapid-Biomed birdcage coil (6.9cm diameter) in active decoupling mode. Spectra were obtained using PRESS sequence (TR/TE=5000/20ms, excited voxel size 1.5x1.5x1.5mm3, depth of 0.9mm, 256 accumulations, experiment duration 20mn). Observations and comparisons were focused on Choline (Cho), Nacetyl-aspartate (NAA), lactate doublet (Lac) and Creatine (Cr) with concentrations of 25, 50 and 100mM in buffered aqueous solutions. Knowing that these cerebral metabolites are not existing biologically with high concentration inside the brain compared to the buffered aqueous solutions. According to the literature review, the concentrations are ranging in the interval of [0.9, 2.5mmol/kg] for Cho, [7, 16mmol/kg] for NAA and [5.01, 10.6mmol/kg] for Cr.8-11 The Figure 1 shows a spectrum obtained using micro-coil. All our Spectra were quantified using Bruker Top spin and via jMRUI software and AMARES method12 (Figure 3). LODs determined in all cases from signal-to-noise measurements were compared according to both of coil strategies and statistically processed. FOG values are reported in Figure 4 for Cho, NAA, Lac and Cr using different concentration 25, 50 and 100 mM. They show that the micro-coil approach can provide a superlative sensitivity and a lower LOD than the surface-coil. Considering significant metabolites as Cho, NAA, Lac and Cr. The FOG is between 2 and 2.4. of the gain for Cr, Cho, NAA and Lac with concentration 25(white), 50(grey), 100(black) mM. In the proposed experiments, the two kinds of coils were used with identical electrical configurations (tuning, matching, preamp bounding). Consequently, the values of sensitivity, LOD and FOG show the advantage of the micro-coil in spite of its smaller sensitive volume to detect and to analyze a small metabolites concentration.
In vivo animal NMR studies using implantable micro-coil and Brain tissue response to chronically implantation
The possibility of using a new generation of micro-coils was brought down in a recent study. It has demonstrated a prosperous potential of enhancement in terms of SNR, spatial resolution, and limits of detection (LOD), compared to the surface-coil. The originality of the present work is to show how to realize brain implantable NMR probehead, tolerated by rats for several weeks. To the best of our knowledge, no in vivo study has been already established with such sensing methodology. Reinforced by acquisition and signal processing, the aim is to push to the limits of in vivo detection and an ultra-localization technique should result from this approach. For in vivo implantation, we have conducted the experiment on rats Wistar (230-300g). The animals handling was carried out according to recommendations of the local ethics committee.13 The micro-coil, "needle" part only, was introduced into the brain by stereotaxy (David-Kopf). First results demonstrate that micro-coil positioning in intra-cerebral structure is well reproducible under stereotaxic conditions. With chronic implantation, the rats survived up to 4 months. Consequently, the biocompatibility aspect is also validated. Axial MRI slices,Figure 5A, allow measurements of the lesion dimensions, compared then with the micro-coil size and to stereotaxic coordinates determined thanks to the rat atlas14 and histological control, Figure 5B. The chosen location coordinates (artery basilar, cerebellum) are visible by MRI. The obtained in vivo water proton spectrum opens cerebral metabolites measurements capabilities as shown in Figure 6.
An ultra-localization technique will result from the development of this planar microcoil, due to its high spatial resolution primarily when focusing on a specific ROI. Comparing to the surface coil, the planar microcoil have shown a lower limits of detection (LOD) by an average FOG of 2.5 and higher SNR despite its small sensitive volume of detection. The LOD estimated by MRS measurements are in good concordance with analytical values given in literature. These preliminary in-vitro results enabled us to open largely the biomedical applications based on cerebral metabolism investigation by experiments in small animals. Subsequently the proposed “needle-coil” concept implies a breakthrough in biomedical research towards NMR micro-detector used for brain exploration for early diagnosis and treatment follow up. Future work will focus on the design of a Low-noise Conditioning system to amplify furthermore the NMR Signal in order to detect lower LOD of cerebral metabolites. The microcoil approach is a newborn tool that could be exploited in several biomedical applications as Alzheimer’s disease & Brain Tumors Diagnosis.