Faradaic and non-faradaic electrochemical impedance spectroscopy as transduction techniques for sensing applications

Ricardo Adriano Dorledo de Faria,1,2 Luiz Guilherme Dias Heneine,3 Tulio Matencio,4 Younes Messaddeq2,5

1Department of Chemical Engineering, Universidade Federal de Minas Gerais, Brazil
2Universite Laval, Canada
3Department of Applied Immunology, Fundacao Ezequiel Dias, Brazil
4Department of Chemistry, Universidade Federal de Minas Gerais, Brazil
5Institute of Chemistry, UNESP, Brazil

Correspondence: Ricardo Adriano Dorledo de Faria, Center for Optics, Photonics and Lasers (COPL), Pavillon d’optiquephotonique, Université Laval, 2375 rue de la Terrasse, Quebec, Canada, Tel (418) 656 2131 (Ext 407184)

Received: February 17, 2019 | Published: February 28, 2019

Citation: Faria RAD, Heneine LGD, Matencio T, et al. Faradaic and non-faradaic electrochemical impedance spectroscopy as transduction techniques for sensing applications. Int J Biosen Bioelectron. 2019;5(1):29-31. DOI: 10.15406/ijbsbe.2019.05.00148

Download PDF

Abstract

Electrochemical Impedance Spectroscopy (EIS) stands out as a powerful technique for application in electroanalytical devices. Especially due to the possibility of label-free performance, use of miniaturized systems, in situ measurements and the relative low cost, impedimetric sensors have attracted particular attention in the recognition of analytes of medical, environmental and industrial interests. Depending on the presence or absence of redox species in the electrode or in the electrolyte, the technique can be categorized as faradaic or non-faradaic EIS respectively. The choice of the most suitable sensing method relies mainly on the expected application. Despite the non-faradaic methods presents the advantages of the application in point-of-care devices due to the possibility of miniaturization of the electrodes and the absence of redox couple, the faradaic sensors tends to be more sensitive.

Keywords: electrochemical impedance spectroscopy, sensor, faradaic EIS, non-faradaic EIS

Introduction

Electrochemical techniques have been widely employed to detect and/or quantify molecules of medical, environmental and industrial interests. Among traditional techniques such as potentiometry, amperometry and conductometry, the Electrochemical Impedance Spectroscopy (EIS) has gained great attention as an analytical tool for a broad range of analytes, mainly because it can be performed as a non-destructive method and it enables in situ measurements.1,2 Moreover, EIS exempt the use of special reagents, which makes it suitable for label-free applications, which significantly diminishes its cost in relation to other techniques and makes it easier to perform.3 In common, all the electrochemical sensors consist on relating a measurable output electrical signal to the changes in the structure (usually the surface) of an electrode because of the recognition of a target analyte. The in-depth understanding of the mechanisms and fundaments involving the sensing phenomena is a challenge in electroanalytical techniques and can lead to the improvement of the main desirable characteristics of a sensor: the selectivity, stability and sensibility.4 In this sense, this minireview presents the main aspects regarding the EIS as a technique for application in sensors based on its two categories of performance: the Faradaic and non-faradaic EIS.

EIS theory

The EIS basis lies on the development of multiple microscopic processes when an electrode/electrolyte system suffers from an electrical stimulus. This small perturbation can provoke the transfer of electrons in electronic conductors and the transport of charged species from the electrode to the electrolyte (and vice-versa) due to the occurrence of oxidation and reduction reactions. Accordingly, the resultant flow rate of electrons and charges is a function of the resistance of both electrode and electrolyte as well as it depends on the reaction rates that take place on the interface.5 Basically, a sinusoidal AC potential (potenciostatic EIS) or current (galvanostatic EIS) is applied to an electrochemical system under study and the resultant AC current ou potential is measured as a frequency-dependent parameter. The ratio between the applied potential E(ω,t), which can also be superimposed by a constant DC potential, and the current I(ω,t) is the impedance Z(ω,t). Herein, ω is the angular frequency (ω =2πf, where f is the variable frequency), t is the time, |Z| is the magnitude of impedance, Z’ is the real impedance, Z” is the imaginary impedance, in which j is the imaginary number √-1, and φ MathType@MTEF@5@5@+= feaagKart1ev2aaatCvAUfeBSjuyZL2yd9gzLbvyNv2CaerbuLwBLn hiov2DGi1BTfMBaeXatLxBI9gBaerbd9wDYLwzYbItLDharqqtubsr 4rNCHbGeaGqiVu0Je9sqqrpepC0xbbL8F4rqqrFfpeea0xe9Lq=Jc9 vqaqpepm0xbba9pwe9Q8fs0=yqaqpepae9pg0FirpepeKkFr0xfr=x fr=xb9adbaqaaeGaciGaaiaabeqaamaabaabaaGcbaqcLbsacqaHgp GAaaa@3842@ is the angle phase between the potential and current signals.6

Z(ω,t)= E(ω,t) I(ω,t) =| Z(ω,t) | e jφ =| Z(ω,t) |(cosφ+jsinφ) Z(ω,t)=Z'(ω)+jZ"(ω) MathType@MTEF@5@5@+= feaagKart1ev2aaatCvAUfeBSjuyZL2yd9gzLbvyNv2CaerbuLwBLn hiov2DGi1BTfMBaeXatLxBI9gBaerbd9wDYLwzYbItLDharqqtubsr 4rNCHbGeaGqiVu0Je9sqqrpepC0xbbL8F4rqqrFfpeea0xe9Lq=Jc9 vqaqpepm0xbba9pwe9Q8fs0=yqaqpepae9pg0FirpepeKkFr0xfr=x fr=xb9adbaqaaeGaciGaaiaabeqaamaabaabaaGceaqabeaajugibi aadQfacaGGOaGaeqyYdCNaaiilaiaadshacaGGPaGaeyypa0tcfa4a aSaaaKGbagaajugibiaadweacaGGOaGaeqyYdCNaaiilaiaadshaca GGPaaajyaGbaqcLbsacaWGjbGaaiikaiabeM8a3jaacYcacaWG0bGa aiykaaaacqGH9aqpjuaGdaabdaqcgayaaKqzGeGaamOwaiaacIcacq aHjpWDcaGGSaGaamiDaiaacMcaaKGbakaawEa7caGLiWoajugibiaa dwgajuaGdaahaaqcgayabeaajugWaiaadQgacqaHgpGAaaqcLbsacq GH9aqpjuaGdaabdaqcgayaaKqzGeGaamOwaiaacIcacqaHjpWDcaGG SaGaamiDaiaacMcaaKGbakaawEa7caGLiWoajugibiaacIcaciGGJb Gaai4BaiaacohacqaHgpGAcqGHRaWkcaWGQbGaci4CaiaacMgacaGG UbGaeqOXdOMaaiykaaGcbaqcLbsacaWGAbGaaiikaiabeM8a3jaacY cacaWG0bGaaiykaiabg2da9iaadQfacaGGNaGaaiikaiabeM8a3jaa cMcacqGHRaWkcaWGQbGaamOwaiaackcacaGGOaGaeqyYdCNaaiykaa aaaa@89EA@

Typically, impedance data is represented as a function of Z’ and Z” (Nyquist diagram) or as a function of the frequency, |Z| and/or φ MathType@MTEF@5@5@+= feaagKart1ev2aaatCvAUfeBSjuyZL2yd9gzLbvyNv2CaerbuLwBLn hiov2DGi1BTfMBaeXatLxBI9gBaerbd9wDYLwzYbItLDharqqtubsr 4rNCHbGeaGqiVu0Je9sqqrpepC0xbbL8F4rqqrFfpeea0xe9Lq=Jc9 vqaqpepm0xbba9pwe9Q8fs0=yqaqpepae9pg0FirpepeKkFr0xfr=x fr=xb9adbaqaaeGaciGaaiaabeqaamaabaabaaGcbaqcLbsacqaHgp GAaaa@3842@ (Bode diagram). By exploiting these graphical representations in certain frequency ranges, it is possible to assess some interfacial characteristics of an electrode as well as the conductive features of the electrolyte. However, due to the complex composition of the global impedance from an electrochemical cell, the EIS data is commonly modeled to an electrical equivalent circuit in order to further analyze the inductive, capacitive and resistive components.8 EIS comprises two groups: the faradaic and non-Faradaic EIS. The use of one or the other terminology is associated to the occurrence of electrical charge transfer across an interface due to the presence or absence of redox species.3 Based mainly on the characteristics of the transducer substrate, the electrolyte and the application of interest, one can elect the most appropriate technique to develop an electrochemical sensor.

Faradaic sensors

In faradaic EIS, reduction and/or oxidation reactions among electroactive species take place at the electrode leading to the generation of an electrical current3 Thus, the faradaic sensors require the presence of redox probes and the application of DC conditions to promote the development of the electrochemical reactions. This type of EIS receives this name because the generated current obeys the Ohm’s Law, which means that the faradaic current is related to the number of electrons transferred in the redox reactions (n), to the Faraday constant (F), the surface area of the electrode (A) and the flux of electroactive molecules at the interface boundary (j). 9

I=nFAj MathType@MTEF@5@5@+= feaagKart1ev2aaatCvAUfeBSjuyZL2yd9gzLbvyNv2CaerbuLwBLn hiov2DGi1BTfMBaeXatLxBI9gBaerbd9wDYLwzYbItLDharqqtubsr 4rNCHbGeaGqiVu0Je9sqqrpepC0xbbL8F4rqqrFfpeea0xe9Lq=Jc9 vqaqpepm0xbba9pwe9Q8fs0=yqaqpepae9pg0FirpepeKkFr0xfr=x fr=xb9adbaqaaeGaciGaaiaabeqaamaabaabaaGcbaqcLbsacaWGjb Gaeyypa0JaamOBaiaadAeacaWGbbGaamOAaaaa@3BCC@

By means of the current generated from the redox reactions, the faradaic sensors can achieve higher sensitivity in comparison to the non-faradaic devices.10 To model the appropriate equivalent circuit associated to faradaic sensors, besides the electrolyte resistance and the capacitance (that is usually replaced by a constant phase element – CPE), it must take in account the effect of a charge transfer resistance (Rct) and a Warburg impedance (Zw). The Rct is a consequence of the potential generated by the oxidative and reductive phenomena at the electrode as well as of the repulsion or stearic effect caused by the charged species that arrives to the conductive surface.3 Accordingly, the Zw results from the diffusion of the electroactive species from the bulk solution to the electrode. Graphically, Rct can be determined by the diameter of a semicircle in the Nyquist plot and Zw can be deducted from a 45° inclined straight line at low frequencies11 Despite the reduction and oxidation reactions are often associated to the presence of a redox couple in the electrolyte (Fe(CN)63-/4- is commonly employed as a redox pair in the literature), the redox phenomena can also occur in the structure of the electrode. For example, some conductive polymers such as polyaniline, polypirrole and polythiophene possess the ability of conduct charge along their chain due to doping mechanisms.12

Non-faradaic sensors

Contrary to the faradaic EIS, the non-faradaic method does not demand the use of redox couples and, consequently, no reference electrode is necessary, because no DC potential is required10 These features make this kind of sensors more amenable to miniaturization and to online/real-time applications. In this kind of technique, by charging and discharging the double-layer capacitance, it is possible to assess the dielectric changes on the electrode. Thus, the impedance of a non-faradaic sensor arises from the effect caused by the insulating characteristics of the target analyte bond to the conductive substrate.13 Typically, the most valuable impedance parameter in this case is the double layer capacitance and it is only limited by the nature of the charge carriers and their concentration in the boundary layer of the transducer substrate.3 Nonetheless, the |Z| and φ MathType@MTEF@5@5@+= feaagKart1ev2aqatCvAUfeBSjuyZL2yd9gzLbvyNv2CaerbdfwBIj xAHbstHrhAaeXatLxBI9gBaerbd9wDYLwzYbItLDharuavP1wzZbIt LDhis9wBH5garqqtubsr4rNCHbGeaGqiVu0Je9sqqrpepC0xbbL8F4 rqqrFfpeea0xe9Lq=Jc9vqaqpepm0xbba9pwe9Q8fs0=yqaqpepae9 pg0FirpepeKkFr0xfr=xfr=xb9adbaqaaeGaciGaaiaabeqaamaaea qbaaGcbaqcLbuacqaHgpGAaaa@3CD6@ components are adequate transducer parameters.14 In this case, the equivalent circuit consists of a combination of resistors and capacitors. The double layer capacitance (C) is described by the Helmholtz model as a function of the free space permittivity (ε0), the relative solution permittivity (εr), the distance (d) of the Helmholtz layer and the surface area of the electrode (A)15

C= ε 0 ε r d A MathType@MTEF@5@5@+= feaagKart1ev2aaatCvAUfeBSjuyZL2yd9gzLbvyNv2CaerbuLwBLn hiov2DGi1BTfMBaeXatLxBI9gBaerbd9wDYLwzYbItLDharqqtubsr 4rNCHbGeaGqiVu0Je9sqqrpepC0xbbL8F4rqqrFfpeea0xe9Lq=Jc9 vqaqpepm0xbba9pwe9Q8fs0=yqaqpepae9pg0FirpepeKkFr0xfr=x fr=xb9adbaqaaeGaciGaaiaabeqaamaabaabaaGcbaqcLbsacaWGdb Gaeyypa0tcfa4aaSaaaOqaaKqzGeGaeqyTdu2cdaWgaaqaaKqzadGa aGimaaWcbeaajugibiabew7aLLqbaoaaBaaaleaajugWaiaadkhaaS qabaaakeaajugibiaadsgaaaGaamyqaaaa@44B8@

Conclusion

Both faradaic and non-faradaic EIS present relevant advantages as transducer techniques in electrochemical sensors for various applications. In the faradaic mode, the charge transfer resistance is the impedance component correlated to the concentration of an analyte of interest and it is a result of the current generated from the reduction and oxidation reactions among electroactive species. On the other hand, the capacitance is generally the most important transducer term to interpret the recognition of a certain analyte in non-faradaic EIS. Despite less sensitive, the non-Faradaic sensor is a promising candidate for real-time applications because it does not require the presence of redox couples.

Acknowledgements

We thank to the Natural Sciences and Engineering Research Council of Canada (CRSNG) and the Canada Excellence Research Chair in Photonics Innovations (CERCP).

Conflict of interest

Author declares that there is no conflict of interest.

References

  1. Bahadir EB, Sezgintürk MK. A review on impedimetric biosensors. Artificial Cells. Nanomed Biotech. 2016;44(1):248–262.
  2. Park JS, Kim HJ, Lee JH, et al. Amyloid beta detection by faradaic electrochemical impedance spectroscopy using interdigitated microelectrodes. Sensors. 2018;18(2):426.
  3. Daniels JS, Pourmand N. Label-free impedance biosensors: opportunities and challenges. Electroanal. 2007; 19(12):1239-1257.
  4. Pejcic B, Marco R. Impedance spectroscopy: over 35 years of electrochemical sensor optimization. Electrochimica Acta. 2006; 5217–6229.
  5. Macdonald JR, Johnson WB. Fundamentals of Impedance Spectroscopy. In: Barsoukov E, Macdonald JR. Impedance Spectroscopy – theory, experiment and applications. Hoboken, NJ: Willey-Interscience; 2018:1–20.
  6. Sacco A. Electrochemical impedance spectroscopy as a tool to investigate the electroreduction of carbon dioxide: a short review. J CO2 Util. 2018;27:22–31.
  7. Sacco A. Electrochemical impedance spectroscopy: fundamentals and application in dye-sensitized solar cells. Renew Sustain Energy. 2017; 79:814–829.
  8. Ates M. Review study of electrochemical impedance spectroscopy and equivalent electrical circuits of conducting polymers on carbon surfaces. Prog Org Coat. 2011; 71:1–10.
  9. Hammond JL, Formisano N, Estrela P, et al. Electrochemical biosensors and nanobiosensors. Essays Biochem. 2016;60(1):69–80.
  10. Park JS, Kim HJ, Lee JH, et al. Amyloid Beta Detection by Faradaic Electrochemical Impedance Spectroscopy Using Interdigitated Microelectrodes. Sensors. 2018;18:1–11.
  11. Barai A, Chouchelamane GH, Guo Y, et al. A study on the impact of lithium-ion cell relaxation on electrochemical impedance spectroscopy. J Power Sources. 2015;280:74–80.
  12. Jr AR, Soto AMG, Mello SV, et al. An electronic tongue using polypyrrole and polyaniline. Synthetic Metals. 2003;132:109–116.
  13. Lin SP, Vinzons LU, Kang YS, et al. Non-Faradaic Electrical Impedimetric Investigation of the interfacial effects of neuronal cell growth and differentiation on silicon nanowire transistors. ACS Appl Mat Interf 2015;7:9866–9878.
  14. Cecchetto J, Fernandes FCB, Lopes R, et al. The capacitive sensing of NS1 Flavivirus biomarker. Bios Bioelectron. 2017;87:949–956.
  15. Kang J, Wen J, Jayaram SH, et al. Development of an equivalent circuit model for electrochemical double layer capacitors (EDLCs) with distinct electrolytes. Electrochimica Acta. 2014;115:587–598.
Creative Commons Attribution License

©2019 Faria, et al. This is an open access article distributed under the terms of the, which permits unrestricted use, distribution, and build upon your work non-commercially.