Review Article Volume 7 Issue 1
Department of Endocrinology & Faculty of Medicine, CK Birla Group of Hospitals, Director of Endocrinology, SCB Diabetes & Hormone Research Foundation, India
Correspondence: Sujit Bhattacharya, Department of Endocrinology & Faculty of Medicine, CK Birla Group of Hospitals, Director of Endocrinology, SCB Diabetes & Hormone Research Foundation, Gajraj Signature, 5A Sadananda Road, Kolkata 700026, West Bengal, India, Tel +91 (33) 40063658
Received: February 28, 2018 | Published: January 4, 2019
Citation: Bhattacharya S. Navigating the wheel of glucose control: where are we now? Endocrinol Metab Int J. 2019;7(1):4-9. DOI: 10.15406/emij.2019.07.00234
Strict glucose control has been shown to reduce microvascular complications in Diabetes. Navigating the wheel of glucose control is not as simple as it may appear. We have relied upon estimation of blood glucose or the HbA1c which reflects an average control over the past 3 months. Other methods like Serum Fructosamine and 1, 5 Anhydroglucitol can also detect glucose control over a shorter period of time. However, the major limitation is that they cannot predict short term glucose fluctuations or recognise hypoglycaemia. Improvement in technology has led to introduction of various forms of continuous glucose monitoring. Real time continuous monitoring within a closed loop has led to the development of the dream reality, the artificial pancreas. In this review, we will discuss the above methods and their limitations.
Keywords: glucose monitoring, HbA1c, fructosamine, continuous glucose monitoring, ambulatory glucose profile
It is well established that intensive glucose control leads to improved incidence of micro vascular outcomes in both Type1 and Type 2 Diabetes.1,2 Guidelines have already laid down the glycaemic targets to improve endpoints. However, the most important issue of concern is the increased incidence of hypoglycaemia as we try to achieve stricter goals.1,2 We routinely monitor capillary blood glucose and HbA1c for glucose control while newer indices like estimated average glucose (eAG) and MAG (Mean Average glucose) has gained popularity. In certain circumstances, Fructosamine and 1, 5 AnhydroGlucitol can be used but it reflects glucose control over a very short period of time. Mean Amplitude of glycaemic excursion (MAGE)3 and Glycaemic variability (GV) are looked at mostly in research settings and clinical relevance is still doubtful.4−28
Continuous glucose monitoring is helpful in certain circumstances especially where intensive monitoring is required and study daily variations over a timed interval. Recently, another simpler version of continuous monitoring can be done using an Ambulatory Glucose profile. The last two methods have the advantage of not only identifying the glucose peaks but also the troughs which are clinically important in routine care. The efficacy of these intensive regimens requires an accurate method to estimate the degree of glucose control which can be achieved.
Navigating the wheel of glucose control
Comprehensive glucose monitoring is not as simple as it may appear. We have been gluco-centric and focussed on reaching the target. Though well recognized, we have ignored the troughs of glucose which could be equally if not more dangerous and life-threatening and seriously impair the quality of life.
The ideal technique should have the following characteristics:
The indices for defining glycaemic control:
Figure 1 Visualization of glucose variability. Solid line: a given excursion. Dashed line: higher glucose variability due to a higher frequency of oscillation. Dotted line: higher glucose variability due to larger amplitude. Note that the mean and area under the curve are identical in the three situations.
Estimation of blood glucose
Glucose can attach to many proteins via a nonenzymatic, posttranslational process.6 This occurs in two stages.
The large majority of commercially available assays of glycated hemoglobin only measures the stable (irreversible) final product ketoamine and do not measure the labile (reversible) aldimine fraction. Therefore, they reflect long-term (two to three months) average glycemia, relatively unaffected by recent acute fluctuations in glucose levels.7,8
Glycated haemoglobin
The most widely used clinical test is measurement of glycated hemoglobin (also called A1C, hemoglobin A1C, glycohemoglobin, or HbA1C). Red cells are freely permeable to glucose. As a result, glucose becomes irreversibly attached to hemoglobin at a rate dependent upon the prevailing glucose concentration. Approximately 1 percent of red cells are destroyed and a similar percentage replenished every day. Thus, the average amount of A1C changes in a dynamic way and indicates the mean blood glucose concentration over the lifespan of the red cell.7,8 Although the A1C reflects mean blood glucose over the entire 120-day lifespan of the red blood cell, it correlates best with mean blood glucose over the previous 8 to 12 weeks. This strong correlation has been demonstrated in several studies including the DCCT that have calculated average glucose on the basis of frequently measured, usually capillary, glucose levels.7,9,10
A large international study (A1C-Derived Average Glucose or ADAG study) calculated average glucose levels in appx 500 subjects (Type1, Type 2 and normal controls) using a combination of CGM and self-monitoring of blood glucose testing, and established a reliable regression equation that can be used to translate A1C results into an estimated average glucose value.11
Studies suggest that A1C values may also be helpful in the diagnosis of impaired glucose tolerance or overt diabetes mellitus, being simpler to perform and to repeat than the oral glucose tolerance test.
Assay problems—In the past, the results of the DCCT could not be extrapolated widely, because of differences in methodology and a lack of standardization among laboratories.9 A new reference method has been established that will provide for even more reliable worldwide standardization of all A1C assays.12 With this new reference system, A1C results are reported globally in SI (Systeme International) units (mmol/mol) and derived The National Glycohemoglobin Standardization Program (NGSP) units (the same values as reported currently as percent of total hemoglobin) using a master equation. An estimated average glucose (eAG), calculated from the A1C result and based upon the results of the ADAG study described above, will be included in the report The eAG (mg/dL or mmol/L) is a more relevant term for patients who self-monitor blood glucose.11 If there are differences in the eAG derived from the A1C result and the meter-calculated average glucose, further exploration is required. For example, if the eAG is higher than the patient's meter-calculated average glucose, it is possible that fingerstick testing is not being performed at times when the blood glucose is highest, eg, after meals. If the eAG is lower than the meter average, a patient may be having undetected periods of low blood glucose, most often nocturnal hypoglycemia. In such cases, the timing of fingerstick blood glucose monitoring requires adjustment.
Sources of error—Although the international standardization of the A1C assay has decreased potential technical errors in interpreting A1C results, there are other biological and patient-specific factors that may cause misleading results:13
Racial variation — Several studies have shown that A1C concentrations are higher in some ethnic groups (African American, Hispanic, Asian) than in white persons with similar plasma glucose concentrations.17,18 If differences in A1C between races do exist, the differences appear to be small and have not been shown to significantly modify the association between A1C and cardiovascular outcomes,18 retinopathy,19,20 or nephropathy.21 In a cross-sectional study using the National Health and Nutrition Examination Survey (NHANES 2005 to 2008), the prevalence of retinopathy actually increased at a lower A1C level in blacks compared with whites (5.5 to 5.9 versus 6.0 to 6.4 percent).21 Potential mechanisms include genetic differences in intracellular glycation, glucose transport across the erythrocyte membrane, or in erythrocyte survival.22,23
Fructosamine
Many proteins other than hemoglobin undergo nonenzymatic glycation,24 to form advanced glycosylation end products, which may play a direct role in the development of diabetic microvascular complications25 Fructosamine is one ketoamine which can be measured in the laboratory.24 Some of the assays are cheaper and easier to perform than the A1C assay and in general has a good correlation with A1C values.25 There are, however, several potential problems with the use of serum fructosamine measurements:
Intraday variations—The range over which blood glucose levels fluctuate vary within a day has been used to define a variety of "unstable diabetes" [4 and is a useful approximation of 5 to 7 point monitoring for several days. While unstable within-day glucose variation may increase the risk of hypoglycaemia, but there is no correlation with the development of microvascular complications.28
1, 5-anhydroglucitol
1, 5-anhydroglucitol (1,5-AG) is a naturally occurring dietary polyol which is normally filtered and completely reabsorbed by the kidneys leading to stable serum concentrations. However, renal reabsorption of 1,5-AG is competitively inhibited by glucose. Within 24 hours of a rise in serum glucose to beyond the renal threshold for glucose excretion i.e >180 mg/dL (10 mmol/L), serum 1,5-AG concentrations fall as urinary losses increase.29−31 Thus, serum 1,5-AG measurements reflect blood glucose values over the past 24 hours and provide information of daily glucose variations, whereas A1C and fructosamine measurements reflect values over two to three months and one to two weeks, respectively.
Moreover, serum 1,5-AG may reflect postprandial glycaemic excursions better than A1C.31 However, this test is not readily available and neither there is any data to suggest that complementary measurement of 1,5-AG (GlycoMark assay) improves glycaemic control or complications of diabetes more than measurement of A1C alone. So, we do not typically measure 1,5-AG concentrations and even with In patients who have high A1C values despite excellent premeal glucose levels, we prefer to test postprandial blood glucose one to two hours after meals to adjust medications.
Day to day variations—The extent to which blood glucose concentrations vary at the same time each day is another useful measure of overall glycaemic control.5 For patients who test their own blood glucose at the same times every day, this variability can be evaluated simply by scanning down columns of blood glucose measurements over several days. The usual explanation for large daily fluctuations is an erratic lifestyle in terms of eating or exercise habits. These problems should be corrected before increasing the insulin dose. The use of continuous glucose monitoring (CGM) devices can make the identification of within-day and between-day variation easier.
Continuous glucose monitoring
Continuous glucose monitoring (CGM) is a novel method for continuously monitoring blood glucose levels. The earlier CGM monitors did not provide real-time blood glucose data. These meters were downloaded in a clinic, where glucose trends were retrospectively analysed, evaluated and treatment decisions made. Real-time CGM (rt-CGM) systems provide immediate visual feedback about blood glucose levels to the patient. Both non–rt-CGM and rt-CGM methods provide visual graphs of blood glucose results over the period of time. In particular, CGM has successfully been used in clinical trials aimed at reducing A1C in children and adults with type 1 diabetes in which the subjects could make immediate adjustments to insulin dosages. Few have included CGM as a method to improve glucose control in patients with type 2 diabetes (T2DM).32−38
In a study examining the clinical effectiveness of continuous glucose sensing, those with real-time access to the visual display had a reduction in time spent in low and high glucose numbers and increased the amount of time within their target blood glucose levels.32,38
There is growing support for use of CGM as a novel method for facilitating behaviour change in people with T2DMStudies suggests that non–rt-CGM and rt-CGM may be efficacious interventions to improve healthy behaviours and decrease A1C; however, the studies did not address the subjective experiences related to wearing the CGM. Most recently, it has been reported that it is important to consider the psychosocial factors associated with CGM.38,39
Findings from another study suggest that use of CGM in women with T2DM is well tolerated and offers a vivid picture of blood glucose patterns that are ordinarily missed in routine diabetes management. Overall, the women were not overly burdened by the physical and logistical requirements associated with wearing the CGM. Most of the complaints were related to technological and personal aspects of CGM, which should be minimized with newer CGM technologies. For a small subset of the women, the experience elicited negative emotions diabetes emotional distress.32,38
In conclusion, continuous glucose monitoring offers patients with T2DM and their health care practitioners a new tool for assessing the effects of real-life events on blood glucose levels. Future research may determine that the use of CGM in the clinical setting may be more effective than routine use of SBGM at home for improving diabetes health outcomes (Figure 2).
Ambulatory glucose profile
Current glucose monitoring tools have important shortcomings/limitations.40−50
For the broad population of people living with diabetes, improved methods for assessing glucose values is needed.
HbA1c: Provides a 90-day average and considered the Standard of Care, but
Inconsistency with testing blood glucose levels inadequate time to detect patterns in handwritten logbooks during short office visits. CGM: Provides a lot of glucose data, but
AGP is constructed from a modal day plot, which collapses glucose data over several days or weeks. The data is plotted according to time as if they occurred over 24 hours.43,45−46,49−50 It is a visual snapshot of a patient’s typical day, revealing patterns of:
The FreeStyle Libre Pro sensor has a sterile, thin flexible fibre that is inserted just under the skin to continuously measure glucose in the body using interstitial fluid (ISF).43−50
Glucose readings–14 days vs. 30 days
Data suggest that a 12-15 day period of monitoring every 3 months may be needed to optimally assess overall glucose control. (Data on file. Abbott Laboratories Limited).
AGP reveals glucose variability by 2 measures.
The distance between the 10th and 90th percentile curves
A median curve that rises and falls shows glucose instability across the day (Figure 3).
KEY benefits of AGP.44−46,49−50
HCP: The AGP graph clearly illustrates patterns in glucose levels. This enables HCPs to detect potential problems, such as hypoglycemia, hyperglycemia, and glycaemic variability (day to day) and glycaemic stability (intraday) more quickly.
GV helps us predict hypoglycemia: This amount of data is not available with SMBG even with the most engaged patient.
Patients: The information in the AGP graph is presented in a simplified way that facilitates understanding of glucose patterns.
The AGP graph provides a link between day-to-day behavior and glucose levels.
Artificial pancreas:
With the increasing availability of continuous glucose monitoring (CGM) and continuous subcutaneous insulin infusion (CSII), the artificial pancreas (AP) (the commonly accepted term for closed-loop control [CLC] of blood glucose [BG] levels in diabetes) has become a hot area in translational research and industrial development. After a prolonged period of inpatient, clinical research centre trials using cumbersome systems, the field has progressed rapidly over the past 2 years to long-term, free-living studies running AP algorithms on smartphones. Although it is still not a cure, the AP is the most promising advance in the treatment of diabetes at this time.42
Currently it is established that an Artificial Pancreas is a change in the treatment paradigm—not a single-function CGM or CSII device but an adaptable wearable network encompassing the patient in a digital treatment ecosystem. There is now a conviction that a mechanical solution to the problem of maintaining strict control of diabetes without increasing the risk of hypoglycaemia is rapidly progressing to a reality.51
Navigating the wheel of glucose control has never been easy in the dark, deep sea of complications in Diabetes especially in the murky waters of hypoglycaemia. Intensive glucose control has led us to look beyond blood glucose and focus on improved estimation & standardisation of HBA1c. However the major limitation was identifying glucose fluctuations and specially recognising hypoglycaemia. Various forms of continuous glucose monitoring systems have been introduced over the past decade. It seems we are apparently heading towards the shore with the progress of science and digital technology. The dream reality of an Artificial Pancreas is a change in the treatment paradigm. The future holds in simplification of the complex technology and in translation into cost-effective clinical benefits with improved quality of life.